Skip to main content
. Author manuscript; available in PMC: 2022 Mar 30.
Published in final edited form as: Lancet Healthy Longev. 2022 Mar 7;3(3):e135–e142. doi: 10.1016/S2666-7568(22)00005-8

Table 2:

Glucose control, insulin delivery, and usage endpoints in the intention-to-treat analysis population

Closed-loop group (n=36) Sensor-augmented pump therapy group (n=37) Treatment difference (95% CI) p value*

Primary endpoint
Time with glucose 3·9 to 10·0 mmol/L, % 79·9% (7·9) 71·4% (13·2) 8·6 (6·3 to 11·0) <0·0001
Key secondary endpoints
 Time with glucose >10·0 mmol/L, % 16·7% (11·4 to 23·9) 21·4% (16·9 to 36·5) −8·5% (−10·9 to −6·1) <0·0001
 Mean glucose, mmol/L 7·8 (0·7) 8·5 (1·1) −0·7 (−0·9 to −0·5) <0·0001
 HbA1c, mmol/mol 49·3 (7·9) 52·1 (9·2) −2·7 (−4·2 to −1·2) 0·0008
 HbA1c, % 6·7% (0·7%) 6·9% (0·9%) −0·2% (−0·4 to −0·1) 0·0008
 Time with glucose <3·9 mmol/L, % 1·7 (1·3 to 2·4) 1·7 (0·9 to 2·7) −0·1 (−0·3 to 0·2) 0·54
Other secondary endpoints
Time with glucose
 <3·5 mmol/L, % 0·7% (0·5 to 1·1) 0·7% (0·4 to 1·2) 0·0% (−0·2 to 0·1) 0·69
 <3·0 mmol/L, % 0·2% (0·1 to 0·3) 0·2% (0·1 to 0·3) 0·0% (−0·1 to 0·1) 0·69
 >l6·7 mmol/L, % 0·5% (0·2 to 0·8) 0·8% (0·2 to 2·8) −0·7% (−1·0 to −0·3) <0·0001
 Glucose, mmol/L 2·6 (0·5) 2·8 (0·6) −0·2 (−0·3 to −0·1) <0·0001
Glucose coefficient of variation, % 32·5 (4·2) 32·7 (4·5) −0·3 (−1·2 to 0·6) 0·49
Total daily insulin, units per day 46·3 (36·9 to 53·5) 42·9 (36·6 to 53·0) 1·2 (−0·6 to 3·0) 0·20
Total daily basal insulin, units per day 27·7 (18·9 to 32·0) 21·5 (15·9 to 27·0) 4·7 (3·2 to 6·1) <0·0001
Total daily bolus insulin, units per day 20·2 (13·5 to 26·1) 23·4 (17·0 to 29·6) −3·5 (−4·9 to −2·0) <0·0001
Total daily dose, units per kg/day 0·5 (0·5 to 0·6) 0·5 (0·4 to 0·6) 0·0 (0·0 to 0·0) 0·35
Time using continuous glucose monitoring, % 99·7 (99·3–99·9) 99·4 (98·8–99·9) 0·45 (0·06–0·85) 0·026
Time using closed-loop, % 96·7% (95·1–98·0) .. .. ..

Data are mean (SD) or median (IQR). Endpoints calculated from all randomised subjects with at least 168 h of CGM data in at least one period. Glucose data are based on sensor glucose measurements. Treatment difference is calculated as closed loop minus sensor augmented pump therapy. One participant randomised to initial use of sensor-augmented pump therapy did not cross over to closed-loop insulin delivery.

*

Based on a linear mixed model adjusting for period as a fixed effect and site as a random effect.

Tested in hierarchy as listed to control the type 1 error using the fixed-sequence method.

Adjusted for multiple comparisons using Benjamini-Hochberg procedure to control false discovery rate. HbA1c=glycated haemoglobin