TABLE 6.
Comparative outcome measures for type of ventilation for included studies.
| CPAP |
IMV/SIMV |
HFO/HFJV |
|||||||||||
| Authors | Unit of measurement | n | Mean bias | Precision | PE | n | Mean bias | Precision | PE | n | Mean bias | Precision | PE |
| Song (28) | ml/kg/min | 54 | −18.2 | ±124.9 | 57 | 39 | −30.2 | ±94.6 | 69.7 | 8 (HFO) | 38.2 | ±160.1 | 85.5 |
| 8 (HFJ) | −10.9 | ±79 | 76.4 | ||||||||||
| Hsu (33) | ml/kg/min | 37 | −2.8 | ±40.7 | 17.3 | 44 | −1.4 | ±179.1 | 27.4 | 24 (HFO) | −16.2 | ±70.1 | 37.8 |
| Torigoe (30) | ml/min | 37 | 3.6 | ±73.1 | 25 | 10 | −29.6 | ±97.8 | 31.7 | 14 (HFO) | −12 | ±105.9 | 33.2 |
| Van Wyk (34) | ml/kg/min | 335 | −23 | ±84.8 | 78.0 | ||||||||
| Hassan (35) | ml/min | 21 | −21.5 | 39 | −115 | 7 (HFO) | −21.5 | ||||||
**Bias = TEBT – TTE. CPAP, continuous positive airway pressure; SIMV, synchronized intermittent mandatory ventilation; HFOV, high frequency oscillatory ventilation; HFJV, high frequency jet ventilation; LOA, limits of agreement; PE, percentage error; TEBT, thoracic electrical biosensing technology; TTE, transthoracic echocardiography.