TABLE 7.

Accuracy and precision in studies with cardiac shunts.

Authors/Year Unit of measurement n
measurements
Cardiac defect Mean* Bias **
Mean ± SD
LOA (precision) PE
Cardiac surgical intervention
Grollmuss (23) ml 240 TGA switch surgery 3.7 0.27 ±1.06 29.0
Weisz (29) ml 78 PDA ligation 1.25 −0.6 (39%) ±0.75 58.0
Physiological shunt
Blohm (31) ml 12 PDA only 4.1 −0.8 ± 1.73 ±0.98 72.1
63 PDA + PFO −1.1 ± 1.58 ±1.09 56.3
125 PFO only −0.6 ± 1.11 ±0.59 40.2
Van Wyk (34) ml/kg/min 304 PDA 124.4 −28.7 ± 43.7 ±78.1 74.4
Torigoe (30) ml/min 23 PDA ≥ 1.5 mm 317 5.5 ±66.2 21.0
58 PDA < 1.5 mm −36.1 ±119.5 38.6
Hassan (35) ml/min 14 DFLPA > 30 cm/s 295.5 −167 203.8 69

*Mean = (TEBT + TTE)/2. **Bias = TEBT – TTE. Bold data indicates calculated data or data estimated from provided graphs. DFLPA, end diastolic flow in left pulmonary artery; LOA, limits of agreement; PDA, patent ductus arteriosus; PE, percentage error; PFO, patent foramen ovale; SD, standard deviation; TGA, transposition of great arteries.