. 2022 Mar 14;24(3):178–189. doi: 10.1089/dia.2021.0319

Table 3.

Sensor Glucose, Glucose Variability, and Insulin Delivery, During the Run-in Period and Study Phase

  Overall (n = 157)
Adolescents (n = 39)
Adults (n = 118)
Run-ina Studyb P Run-ina Studyb P Run-ina Studyb P
24-h day
 SG, mg/dL 153 ± 16 148 ± 10 <0.001d 162 ± 16 150 ± 8 <0.001 151 ± 15 147 ± 11 <0.001
 CV of SG, % 35.0 ± 5.0 34.2 ± 4.1 0.003 36.5 ± 4.8 35.7 ± 4.1 0.202 34.5 ± 5.0 33.7 ± 3.9 0.008
 Total insulin, units 54.9 ± 25.8 56.7 ± 28.4 0.002d 60.0 ± 16.3 63.1 ± 17.3 0.007 53.2 ± 28.1 54.6 ± 31.0 0.103d
 Total basal insulin, units 25.1 ± 13.6 23.6 ± 13.2 <0.001d 25.2 ± 8.7 24.4 ± 8.4 0.265 25.1 ± 14.9 23.3 ± 14.5 <0.001d
 Total bolus insulin, units 29.8 ± 14.8 33.1 ± 16.8 <0.001d 34.8 ± 10.9 38.7 ± 10.1 0.002 28.1 ± 15.5 31.3 ± 18.2 <0.001d
 Auto Correction, units 0.0 ± 0.4 7.5 ± 6.1 <0.001d 0.0 ± 0.2 8.3 ± 3.3 <0.001 0.0 ± 0.5 7.3 ± 6.7 <0.001d
 Auto Correction, %c 0.2 ± 1.4 22.0 ± 9.3 <0.001d 0.1 ± 0.5 21.6 ± 7.0 <0.001 0.2 ± 1.6 22.1 ± 10.0 <0.001d
Daytime (6 AM–12 AM)
 SG, mg/dL 155 ± 17 151 ± 12 <0.001d 164 ± 17 154 ± 10 <0.001d 151 ± 16 150 ± 12 0.113
 CV of SG, % 35.0 ± 5.1 34.5 ± 4.0 0.157d 36.7 ± 4.9 36.0 ± 4.1 0.261 34.5 ± 5.1 34.0 ± 3.8 0.129
 Total insulin, units 46.4 ± 21.7 47.5 ± 23.6 0.024d 51.4 ± 13.7 53.4 ± 14.2 0.061 44.7 ± 23.6 45.6 ± 25.7 0.250d
 Total basal insulin, units 18.7 ± 10.1 17.2 ± 9.8 <0.001d 18.6 ± 6.6 17.7 ± 6.1 0.092 18.7 ± 11.0 17.0 ± 10.8 <0.001d
 Total bolus insulin, units 27.6 ± 13.8 30.3 ± 15.3 <0.001d 32.7 ± 10.1 35.7 ± 9.3 0.017 26.0 ± 14.5 28.5 ± 16.5 <0.001d
 Auto Correction, units 0.0 ± 0.4 5.9 ± 4.8 <0.001d 0.0 ± 0.2 6.5 ± 2.7 <0.001d 0.0 ± 0.4 5.7 ± 5.4 <0.001d
 Auto Correction, %c 0.2 ± 1.3 18.8 ± 8.7 <0.001d 0.1 ± 0.5 18.5 ± 6.6 <0.001d 0.2 ± 1.5 18.9 ± 9.2 <0.001d
Nighttime (12 AM–6 AM)
 SG, mg/dL 150 ± 19 140 ± 13 <0.001 156 ± 19 140 ± 12 <0.001 148 ± 19 140 ± 14 <0.001
 CV of SG, % 33.1 ± 6.1 30.6 ± 5.2 <0.001d 34.1 ± 5.9 32.0 ± 5.0 0.008 32.7 ± 6.1 30.1 ± 5.2 <0.001d
 Total insulin, units 8.6 ± 4.8 9.3 ± 5.5 <0.001d 8.7 ± 3.5 9.8 ± 3.8 <0.001d 8.5 ± 5.1 9.1 ± 5.9 <0.001d
 Total basal insulin, units 6.5 ± 3.7 6.4 ± 3.7 0.990d 6.6 ± 2.3 6.7 ± 2.3 0.593 6.4 ± 4.0 6.4 ± 4.0 0.672d
 Total bolus insulin, units 2.1 ± 2.1 2.8 ± 2.5 <0.001d 2.1 ± 2.1 3.1 ± 2.2 <0.001 2.1 ± 2.0 2.8 ± 2.6 <0.001d
 Auto Correction, units 0.0 ± 0.1 1.7 ± 1.5 <0.001d 0.0 ± 0.0 1.8 ± 0.9 <0.001d 0.0 ± 0.1 1.6 ± 1.6 <0.001d
 Auto Correction, %c 0.5 ± 3.6 66.1 ± 22.4 <0.001d 0.3 ± 1.3 69.9 ± 20.6 <0.001d 0.5 ± 4.1 64.9 ± 22.9 <0.001d

All values are shown as mean ± SD.

Run-in CGM was ∼14 days and study phase was ∼90 days.

During baseline run-in period, Auto Correction bolus was inadvertently delivered on six systems.

a

Sensor-augmented/integrated pump, PLGM, or Auto Basal use.

b

Auto Basal+Auto Correction use.

c

Factored as a percentage of total bolus insulin.

d

Wilcoxon signed-rank test.

CV, coefficient of variation; SG, sensor glucose.