Skip to main content
. 2000 Mar;44(3):546–550. doi: 10.1128/aac.44.3.546-550.2000

TABLE 1.

Validation parameters for the enzyme immunoassay of AMB in plasma and selected tissues

Parameter (%) Value for:
Plasmaa with AMB at (pg/ml):
Tissueb
1,000 500 200 100 Lung Brain
Intraday CVc 3 6 7 5 26 19
Interday CVd 16 15 17 11 17 9
Accuracye 102 106 92 111 118 82
a

Human plasma spiked with AMB at the concentrations indicated. 

b

Mouse tissues spiked with AMB at the concentration of 20 ng/g. 

c

n = 8 for plasma; n = 4 for tissues. 

d

n = 5 for plasma; n = 4 for tissues. 

e

Recovery of added AMB in the interday experiment.