TABLE 1.
Validation parameters for the enzyme immunoassay of AMB in plasma and selected tissues
| Parameter (%) | Value for:
|
|||||
|---|---|---|---|---|---|---|
| Plasmaa with AMB at (pg/ml):
|
Tissueb
|
|||||
| 1,000 | 500 | 200 | 100 | Lung | Brain | |
| Intraday CVc | 3 | 6 | 7 | 5 | 26 | 19 |
| Interday CVd | 16 | 15 | 17 | 11 | 17 | 9 |
| Accuracye | 102 | 106 | 92 | 111 | 118 | 82 |
Human plasma spiked with AMB at the concentrations indicated.
Mouse tissues spiked with AMB at the concentration of 20 ng/g.
n = 8 for plasma; n = 4 for tissues.
n = 5 for plasma; n = 4 for tissues.
Recovery of added AMB in the interday experiment.