Table 1. Parameters of quadrant-detection adaptive optics scanning light ophthalmoscope system.

System Parameters Value
Imaging Wavelength and Peak Power a 790 nm, 150 µW
Beacon Wavelength and Peak Power a 830 nm, 20 µW
Beam Size at Pupil of the Eye 7.75 mm
Confocal Detection Pinhole Size 1 Airy Disk
Non-Confocal Annulus Outer Diameter 20 Airy Disks
Non-Confocal Annulus Inner Diameter 2 Airy Disks
Field of View ∼1×1°
Horizontal Scanning Rate 15 kHz
Horizontal/Vertical Number of Pixels 768/650
Digitizer Pixel Clock 40 MHz
Frame Rate 16.6 fps
a

The continuous wave light source (CW) was turned ON only in the portions of the scanner raster necessary for imaging, and off elsewhere. Therefore, average power was approximately 2.5 times lower than peak power.