. 2021 Oct 20;157(4):602–607. doi: 10.1093/ajcp/aqab173

TABLE 3.

LumiraDx SARS-CoV-2 Ag Test Performance in Cycle Threshold Subsets as Determined by the cobas 6800 SARS-CoV-2 Test (Roche Diagnostics), TaqPath COVID-19 Combo Kit (Thermo Fisher Scientific), and Either Reverse Transcription–Polymerase Chain Reaction Method

RT-PCR Test Roche Diagnostics + Thermo Fisher Scientific Roche Diagnostics Thermo Fisher Scientific
Ct Value No. PPA (95% CI), % No. PPA (95% CI), % No. PPA (95% CI), %
<33 24 95.8 (79.8-99.3) 8 100 (67.6-100) 16 93.8 (71.7-98.9)
<30 22 100 (85.1-100) 7 100 (64.6-100) 15 100 (79.6-100)
<25 18 100 (82.4-100) 6 100 (61.0-100) 12 100 (75.8-100)

CI, confidence interval; COVID-19, coronavirus disease 2019; Ct, cycle threshold; PPA; positive percent agreement; RT-PCR, reverse transcription–polymerase chain reaction; SARS-CoV-2, severe acute respiratory syndrome coronavirus 2.