Table 1.

Minimal technical requirements for multiparametric prostate MRI according to the PI-RADS v.2.1 guidelines

T 2-weighted imaging
(T2-WI)
Diffusion-weighted imaging
(DWI)
Dynamic contrast-enhanced
(DCE)
Imaging planes Same used for DWI and DCE Same used for T2-WI and DCE Same used for T2-WI and DWI
Slice thickness 3 mm, no gap ≤4 mm, no gap 3 mm, no gap
Field of view 12–20 cm a 16–22 cm 12–20 cm a
In-plane dimension ≤0.7 mm (phase) x ≤ 0.4 mm (frequency) ≤2.5 mm (phase and frequency) ≤2 mm (phase and frequency)
Specific recommendations
T2-WI acquisition Axial plane: either straight axial to the patient or in an oblique axial plane matching the long axis of the prostate - -
At least one additional orthogonal plane (sagittal and/or coronal) - -
3D axial as an adjunct to 2D acquisitions - -
Low b-value - 0 (preferably 50)–100 sec/mm2 -
Intermediate b-value - 800–1000 sec/mm2 -
High b-value -
  • Dedicated (≥1,400 sec/mm2)

  • Synthesised (from other b-values)

-
Temporal resolution - - ≤15 s
Total observation rate - - >2 min
Dose of GBCA - - 0.1 mmol/kg
Injection rate - - 2 - 3cc/s
Fat suppression/subtraction - - Recommended

Legend – T2-WI: T 2-weighted imaging; DWI: diffusion-weighted imaging; DCE: dynamic contrast enhanced; GBCA: Gadolinium-based contrast agent

a

to encompass the entire prostate gland and seminal vesicles