Table 5.

Mean bias between consumer devices and medical devices

TST (min) WASO (min) Light (min) Deep (min) REM (min) SOL (min) NAWK (count) SE (%)
Fitbit − 12.3 a 24.5 − 42.4 39.8 − 11.6 − 11.1 3.1 1.5
Neuroon − 156.5 172.6 − 165.2 20.4 − 19.6 52.6 − 13.6 − 36.3
Neuroon (signal quality = 2) − 6.9 4.4 − 64.7 38.0 11.6 4.5 − 15.5 1.7
Neuroon (signal quality = 0) − 265.1 281.6 − 223.6 5.8 − 62.1 94.7 − 12.9 − 62.7

aBold indicates bias within acceptable range