Table 2.
Summary of acceptance criteria used to assess plate uniformity in intra-plate tests and the determined values for the plate uniformity test carried out on the human DRD2 Gz TRUPATH assay
Criterion | Acceptance range | Observed values for DRD2 Gz TRUPATH assay |
---|---|---|
%CV Max | ≤20% | 2.49 |
%CV EC80 | ≤20% | 3.58 |
%CV EC50 | ≤20% | 4.36 |
%CV Min | ≤20% | 4.06 |
Normalised SD EC80 | ≤20 | 5.39 |
Normalised SD EC50 | ≤20 | 7.77 |
Z prime (max and min) | ≥0.4 | 0.63 |
% coefficient of variation (%CV) and standard deviation (SD). Criterion and acceptance criteria adapted from Iversen et al., 2004.
Data shown in Figure 10.