Skip to main content
. 2022 Apr 1;3(2):101259. doi: 10.1016/j.xpro.2022.101259

Table 2.

Summary of acceptance criteria used to assess plate uniformity in intra-plate tests and the determined values for the plate uniformity test carried out on the human DRD2 Gz TRUPATH assay

Criterion Acceptance range Observed values for DRD2 Gz TRUPATH assay
%CV Max ≤20% 2.49
%CV EC80 ≤20% 3.58
%CV EC50 ≤20% 4.36
%CV Min ≤20% 4.06
Normalised SD EC80 ≤20 5.39
Normalised SD EC50 ≤20 7.77
Z prime (max and min) ≥0.4 0.63

% coefficient of variation (%CV) and standard deviation (SD). Criterion and acceptance criteria adapted from Iversen et al., 2004.

Data shown in Figure 10.