Table 2.
Summary of acceptance criteria used to assess plate uniformity in intra-plate tests and the determined values for the plate uniformity test carried out on the human DRD2 Gz TRUPATH assay
| Criterion | Acceptance range | Observed values for DRD2 Gz TRUPATH assay |
|---|---|---|
| %CV Max | ≤20% | 2.49 |
| %CV EC80 | ≤20% | 3.58 |
| %CV EC50 | ≤20% | 4.36 |
| %CV Min | ≤20% | 4.06 |
| Normalised SD EC80 | ≤20 | 5.39 |
| Normalised SD EC50 | ≤20 | 7.77 |
| Z prime (max and min) | ≥0.4 | 0.63 |
% coefficient of variation (%CV) and standard deviation (SD). Criterion and acceptance criteria adapted from Iversen et al., 2004.
Data shown in Figure 10.