TABLE 2.
Measurements taken in triplicate under standard conditions for different dose rates
Measured doses (cGy) | ||||||
---|---|---|---|---|---|---|
Dose rate (MU/min) | 1st | 2nd | 3rd | StDev (cGy) | Mean dose (cGy) | Relative dose (%) |
20 | 103.74 | 103.84 | 104.02 | 0.14 | 103.87 | 103.6 |
60 | 102.04 | 102.04 | 102.15 | 0.06 | 102.08 | 101.81 |
80 | 101.81 | 101.72 | 101.61 | 0.1 | 101.71 | 101.45 |
100 | 101.40 | 101.48 | 101.44 | 0.04 | 101.44 | 101.18 |
200 | 101.10 | 101.08 | 100.82 | 0.16 | 101 | 100.74 |
300 | 101.16 | 101.09 | 101.06 | 0.05 | 101.1 | 100.84 |
400 | 100.56 | 100.56 | 100.57 | 0.01 | 100.56 | 100.3 |
600 | 100.42 | 100.19 | 100.17 | 0.14 | 100.26 | 100 |