. 2022 Jan 19;23(4):e13525. doi: 10.1002/acm2.13525

TABLE 2.

The mean (± 1 standard deviation (SD)) difference between synthetic computed tomography (CT) and deformed CT‐based patient positioning, as well as synthetic digital reconstructed radiograph (DRR) and deformed DRR, in all directions

Translation (mm) Rotation (°)
Registration Direction mean (± 1SD) (minimum, maximum) mean (±1SD) (minimum, maximum)
3D evaluation using Cone Beam CT Right‐left/pitch 0.1 ± 0.5 (−1.0, 1.0) 0.3 ± 0.5 (−0.2, 1.3)
Posterior‐anterior/yaw 0.4 ± 0.7 (−0.7, 1.9) 0.0 ± 0.3 (−0.5, 0.7)
Inferior‐superior/roll −0.7 ± 0.6 (−2.1, 0.3) 0.0 ± 0.4 (−0.9, 0.7)
2D evaluation using orthogonal projections Right‐left/pitch −0.0 ± 0.7 (−2.5, 1.5) 0.1 ± 0.6 (−1.5, 1.8)
Posterior‐anterior/yaw −0.2 ± 0.7 (−2.5, 1.1) −0.1 ± 0.5 (−1.3, 1.2)
Inferior‐superior/roll −0.3 ± 0.8 (−2.5, 1.3) NaN (NaN, NaN)