TABLE 1.

Cohort 1 failure modes and effects analysis (FMEA) results, showing average O, S, D, and risk priority number (RPN) scores

Process step Potential failure mode Potential cause of failure Occurrence (O) Potential effects of failure Severity (S) Detection methods‐current controls Detectability (D) RPN = (O) (S)(D) Recommended action
Set up with FB DICOM surface Patient within SGRT tolerance but not positioned correctly SGRT tolerance too wide due to incorrect site or tolerances adjusted at linac 4.7 Treatment not delivered accurately 7.2 Immobilization, table tolerances, SSDs, physics pretreatment QA, imaging 2.8 95

Establish protocol: physics review FB versus DIBH AP SSDs per patient and adjust tolerances as needed prior to start date

Checklist: create Aria Encounter to aid in process

Patient within SGRT tolerance but not positioned correctly SGRT fooled: not enough landmarks in ROI 5.0 Treatment not delivered accurately 6.7 Immobilization, table tolerances, SSDs, physics pretreatment QA, imaging 3.3 111 Checklist: create therapist daily Aria Encounter to verify ROI; add to physics weekly encounter to verify ROI
Pull up correct SGRT surface (DICOM or VRT) Use DICOM mistakenly instead of VRT surface Imaging form not updated from DICOM; patient moved to other linac 5.7 Treatment not delivered accurately 5.8 Imaging form, imaging 3.0 99

Computerization: map network drive and train therapists how to move files between linacs

Checklist: create log to track which patients moved and if images taken

Capture new VRT surface if shifts from imaging Patient moves or breathes during process Patient discomfort 5.2 Treatment not delivered accurately 5.3 SGRT, visual inspection of patient, imaging 4.2 115 Establish protocol: change imaging policy to require one therapist to watch patient while the other captures VRT surface; take verification port after all shifts to verify VRT surface
Daily bolus surface captured during DIBH Patient moves or breathes during process Patient discomfort 4.7 Treatment not delivered accurately 5.2 SGRT, visual inspection of patient, imaging 4.0 96 Barrier: investigate using rigid bolus to avoid daily VRT captures
Bolus capture taken with incorrect surface chosen (DIBH or FB) Labels incorrect or confusing 4.0 Treatment not delivered accurately 5.7 Dates and appearances of SGRT surfaces 4.2 94

Labels: standardize labeling and order of appearance in SGRT; standardize plan order on linac console

Checklist: physics to check coincidence of shifts and imaging form changes during weekly QA

Bolus capture taken with incorrect surface chosen (DIBH or FB) Lack of attention 4.7 Treatment not delivered accurately 5.5 Dates and appearances of SGRT surfaces 4.3 111

Barrier: investigate using rigid bolus to avoid daily VRT captures

Checklist: create therapist daily Aria Encounter to verify ROI

Abbreviations: AP, anterior‐posterior; DIBH, deep inspiration breath‐hold; DICOM, surface created from treatment planning CT; FB, free breathing; QA, quality assurance; ROI, region of interest; SGRT, surface‐guided radiation therapy; SSD, source surface distance; VRT, vision RT generated surface.