TABLE 7.
Ratio of absorbed dose to water of modified calibration and Technical Report Series (TRS)‐398 () obtained from parallel‐plate chamber used in this work
Elekta Synergy Platform | Elekta Versa HD | ||||||
---|---|---|---|---|---|---|---|
Type of chamber | Energy (MeV) |
|
Discrepancy (%) |
|
Discrepancy (%) | ||
Exradin A11 | 6 | 1.013 | 1.3 | 1.013 | 1.3 | ||
8 | 1.013 | 1.3 | 1.013 | 1.3 | |||
10 | 1.012 | 1.2 | 1.013 | 1.3 | |||
12 | 1.014 | 1.4 | 1.014 | 1.4 | |||
15 | 1.015 | 1.5 | 1.016 | 1.6 |
Note: Discrepancy (%) = (absorbed doses for modified calibration – absorbed doses for TRS‐398)/(absorbed doses for TRS‐398 dose).