Skip to main content
. 2022 Apr 6;27(7):2353. doi: 10.3390/molecules27072353

Table 1.

Analytical performances and EF values of six estrogens using in-tube SPME-HPLC method.

Analytes LODs (μg/L) LOQs (μg/L) Linear Ranges (μg/L) R2 EF Extraction Repeatability
(n = 3, RSD%) a
Preparation Repeatability
(n = 3, RSD%) b
Intraday Interday
Bisphenol A 0.097 0.323 0.5–10.0 0.9857 20 2.0 4.6 10.4
Estradiol 0.135 0.450 0.5–10.0 0.9886 33 2.1 4.8 8.9
Ethynyl estradiol 0.131 0.437 0.5–10.0 0.9886 28 1.3 3.2 5.2
Estrone 0.079 0.263 0.5–10.0 0.9867 177 0.8 4.5 14.3
Diethylstilbestrol 0.132 0.440 0.5–10.0 0.9794 43 1.7 4.0 2.3
Hexestrol 0.092 0.307 0.5–10.0 0.9910 154 0.9 2.9 9.4

a Extraction repeatability was investigated by extracting 10 μg/L of estrogens three times. b Preparation repeatability was investigated by extracting 10 μg/L of estrogens with three extraction tubes.