Table 4.

Partial validation of hydroxycitric acid (HCA) in GD18 Hsd:Sprague Dawley® SD® rat fetal homogenate

Matrix Concentration Range (ng/g) 60-3000
Coefficient of Determination (r2) > 0.993
LOD a (ng/g) 7.77
LOQ (ng/g) 60.0
Accuracy (RE b%)
Intra-Day c ≤ ±12.3
Intra-Day d ≤ ±11.9
Precision RSDe (%)
Intra-Day c ≤ 14.5
Intra-Day d ≤ 5.5
Method Verification
Accuracy (REb %) 2.6
Precision (RSDe %) 2.8
a

LOD = Limit of Detection; 3 x Standard Deviation for 6 lower limit of quantitation (LOQ) replicates

b

%RE = percent relative error

c

Estimated based on fetal homogenate calibration standards

d

Estimated based on fetal homogenate on low-, mid-, and high-level (90, 900, and 2260 ng/g, respectively) Quality Control (QC) samples

e

%RSD = percent relative deviation.