Table 1.
Three stan dardization stages |
Four flow cytometry standardization categories |
|||
---|---|---|---|---|
Equipment | Antibody panels | Antibodies/staining reagents | Analysis/reporting | |
Fundamental | Different instruments Performance of instrument calibration using CS&T and application settings (BD), FSP (Beckman Coulter), or SpectroFlo (Cytek Aurora) beads Challenge: requires thorough data comparisons across different types of instruments |
Different antibody panel at each site, albeit same marker(s) to define cell types. Check full panel stain on samples relevant to the trial (e.g., subjects with T1Ds) Challenge: different clones can lead to differences in MFI and percentage of positive cells detected |
Perform antibody titrations and optimization of other key reagents. Ensure regents are used within expiry dates Challenge: time- and resource-consuming to check titration of different antibody lots |
Similar gating strategy and the use of appropriate controls. Report all deviations from protocol that may impact analysis Challenge: a rigid analysis may preclude unexpected results |
Desirable | Similar instruments, despite different configurations (e.g., BD LSRFortessa 3 vs. 4 lasers) Back-up instruments at each site Challenges: requires thorough data comparisons across different instruments; some sites will limit and prioritize markers due to lack of number of parameters |
Similar panels across sites that use same markers and clones albeit a slight difference in marker/fluorochrome combination Challenge: different fluorochromes usually affect marker staining sensitivity; must be taken into account |
Marker evaluation across platforms to ensure that populations of interest can be detected by cytometers at each clinical trial site. Challenge: time- and resource-consuming |
Predefined gating strategy and data acceptance criteria along with normalization to controls Challenge: a rigid analysis may preclude unexpected results |
Ideal | Identical instruments with the same configuration Levy-Jennings plots for cross-validating QC across sites & over time Challenge: same instrument can differ in daily performance and sensitivity; some comparisons of data is desirable |
Identical panels with consideration to minimize fluorochrome spillover and spread Challenge: there may be lot-to-lot variability between identical antibodies. Consider purchasing lots in bulk |
Use of lyophilized antibody tubes to reduce or eliminate lot-to-lot variation and pipetting error Perform lot-to-lot crossovers when using wet antibodies Challenge: when several panels are used for a study, lot-to-lot crossover testing can become tedious, expensive, and not feasible |
Use of automatic gating and independent analyses of same samples/data by two different sites. Internal controls to assess technical versus biological variation Challenge: a rigid analysis may preclude unexpected results |
Case study | Desirable Site B: 5-laser BD LSRFortessa Site C: 3-laser BD LSRFortessa Ideal improvement: use of same instrument with identical configuration to allow for the use of identical flow panels |
Desirable Similar five flow panels using same antibody clones Ideal improvement: identical instrument used with identical flow panels |
Desirable Wet antibody titrated with documentation of reagent lot numbers Ideal improvement: validate the use of antibody master mixes and/or identify where possible the use of lyophilized antibody tubes to reduce pipetting |
Ideal Identical gating strategy, data analyzed by site each and all data reviewed by coordinating site Operator and time between sample collection and processing were documented to assess technical variation |
The application of three stages of standardization (fundamental, desirable, and ideal) to four categories (equipment, antibody panel design, antibodies/staining reagents, and analysis/reporting) when designing flow cytometric assays in a clinical trial and when applied to a case study with challenges noted.