Table 2. Radiation dose for the AiCE and AIDR 3D protocols.

Parameter (mean ± SD) All patients BMI <20 (n=7) BMI 20–24.9 (n=37) BMI ≥25 (n=27)
AiCE AIDR 3D Dose reduction (%) AiCE AIDR 3D Dose reduction (%) AiCE AIDR 3D Dose reduction (%) AiCE AIDR 3D Dose reduction (%)
CTDIvol (mGy) 10.8±3.9 19.4±5.6 44.3 6.9±1.1 11.8±2.3 41.5 8.7±2.1 16.8±3.1 48.2 14.7±2.9 25.0±3.5 41.2
ED (mSv) 4.9±2.0 8.7±3.0 43.7 3.1±0.7 5.1±1.4 39.2 3.8±1.0 7.3±1.6 47.9 6.9±2.4 11.6±2.1 40.5
SSDE (mGy) 14.6±4.1 26.5±5.7 44.9 10.5±1.2 18.4±2.7 42.9 12.5±2.4 24.0±3.4 47.9 18.7±2.9 32.0±3.3 41.6

AiCE, advanced intelligent clear-IQ engine; AIDR 3D, adaptive iterative dose reduction 3D; SD, standard deviation; BMI, body mass index; CTDIvol, volume CT dose index; ED, effective dose; SSDE, size-specific dose estimate.