Table 3. The mean image noise and CNR for the AiCE and AIDR 3D protocols.

Parameter All patients BMI <20 (n=7) BMI 20–24.9 (n=37) BMI ≥25 (n=27)
AiCE AIDR 3D P* AiCE AIDR 3D P* AiCE AIDR 3D P* AiCE AIDR 3D P*
Image noise (mean ± SD)
   Erector spinae muscles 11.7±6.3 12.9±6.6 <0.001 9.5±1.7 11.6±3.4 0.028 10.8±6.2 11.7±6.5 0.002 13.6±6.8 14.8±7.0 0.054
   Hepatic parenchyma 9.2±0.9 9.8±0.9 <0.001 8.9±0.8 9.2±0.7 0.391 8.9±0.8 9.5±0.7 <0.001 9.6±1.0 10.4±0.7 <0.001
CNR (mean ± SD)
   Hepatic parenchyma 5.0±1.5 4.3±1.3 <0.001 5.3±1.5 4.3±0.9 0.033 5.3±1.3 4.7±1.3 0.001 4.4±1.5 3.7±1.2 <0.001

*, statistical comparison using t-test. CNR, contrast-to-noise ratio; AiCE, advanced intelligent clear-IQ engine; AIDR 3D, adaptive iterative dose reduction 3D; BMI, body mass index; SD, standard deviation.