Table 2. Radiation doses of different tube current.

mA 20 30 50 70 90 100 120 150
CTDIvol (mGy) 0.13 0.2 0.34 0.47 0.61 0.67 0.81 1.01
DLP (mGy*cm) 4.88 7.32 12.19 17.07 21.95 24.38 29.25 36.57
ED (mSV) 0.06 0.10 0.17 0.23 0.30 0.34 0.41 0.51