Table 2. Technical parameters of different MR sequences.

Parameters Two-echo Dixon, SAT/VAT volume Six-echo Dixon, supraclavicular PDFF Six-echo Dixon, gluteal PDFF
TR, ms 4.0 7.8 8.3
TE1/TE2/ΔTE, ms TE1 =1.32 TE1 =1.24 TE1 =1.34
TE2 =2.60 ΔTE =1.00 ΔTE =1.10
Flip angle 10°
Bandwidth, Hz/pixel 1,004 1,413 1,437
Acquisition matrix size 332×219×44 268×201×93 200×267×40
Reconstruction matrix size 512×512 288×288 320×320
Field of view (FOV), mm3 500×445×220 400×300×140 400×400×60
Acquisition voxel 1.5×2.03×5.0 1.5×1.5×1.5 1.5×1.5×1.5
Reconstructed voxel, mm3 0.98×0.98×5.00 1.389×1.389×1.500 1.25×1.25×1.50
Slices 44 93 40
SENSE reduction factor 2.5x1.0 (AP 2.5) 2.5×1.0 (AP 2.5) 2.5×1.0 (LR 2.5)
Partial Fourier 075×0.80 No No
Respiratory triggering No Yes No
Scan time 10 s 2 min 30 s 46 s

TE1, first echo time; TE2, second echo time; ΔTE, difference between echo times. AP, anterior posterior; FOV, field of view; LR, left right; MR, magnetic resonance; PDFF, proton density fat fraction; SAT, subcutaneous adipose tissue; TR, repetition time; TE, echo time; SENSE, sensitivity encoding; VAT, visceral adipose tissue.