Diagnostic measures of the 6-lead ECG of the device for the detection of AF.

Measure AF vs. stable SR group (n = 214) AF vs. both SR groups including frequent PVCs/PACs (n = 341)
Sensitivity (%), (95% CI) 99.2 (95.4–100) 99.2 (95.4–100)
Specificity (%), (95% CI) 100 (96.2–100) 99.1 (96.8–99.9)
Accuracy (%), (95% CI) 100 (-) 99.1 (97.4–99.8)
LR (+), (95% CI) - 110.07 (27.70–437.41)
LR (-), (95% CI) 0.01 (0.00–0.06) 0.01 (0.00–0.06)

AF, atrial fibrillation; SR, sinus rhythm; PVC, premature ventricular contraction; PAC, premature atrial contraction; LR (+), positive likelihood ratio; LR (-), negative likelihood ratio.