Skip to main content
. 2022 Mar 22;60(4):e02188-21. doi: 10.1128/jcm.02188-21

TABLE 5.

Ana-CIM reproducibility testing by number of tests with zone size off the modea

Isolateb Mode for all tests (mm) No. of tests with zone size (mm) off the mode
<−5 −5 −4 −3 −2 −1 0 +1 +2 +3 +4 +5 >+5
R1 19 4 29 37 10 1
R2 6 78 3
R3 19 2 27 41 11
R4 6 81
R5 6 39 1 41
Kp pos control 6 81
Bf pos control 6 81
Bf neg control 19 3 29 30 19
Total 0 0 0 0 9 85 468 41 1 0 0 0 44
a

Kp pos control, K. pneumoniae ATCC BAA-1705; Bf pos control, B. fragilis WIS-ImiR-001; Bf neg control, B. fragilis ATCC 25285.

b

Each isolate was tested a total of 81 times (3 replicates per day on 3 different testing days at 3 different sites).