| scan type |
selected reaction monitoring
(SRM) |
| source |
heated-electrospray ionization (H-ESI) |
| ion polarity |
positive |
| vaporizer temperature (°C) |
300 |
| ion transfer tube temperature (°C) |
250 |
| sheath gas (arbitrary scale) |
30 |
| auxiliary gas (arbitrary scale) |
10 |
| CID gas (mTorr) |
1.5 |
| dwell time (ms) |
148 |
| RF lens for amodiaquine (V) |
154 |
| RF lens for N-desethylamodiaquine (V) |
160 |
| RF lens for IS (V) |
58 |
| Q1 resolution (FWHM) |
0.7 |
| Q3 resolution (FWHM) |
0.7 |
| collision energy for amodiaquine (V) |
18 |
| collision energy for N-desethylamodiaquine (V) |
17 |
| collision energy for IS (V) |
20 |
| ion transition for amodiaquine (m/z) |
356.2 → 283.1 |
| ion transition for N-desethylamodiaquine (m/z) |
328.2 → 283.1 |
| ion transition for IS (m/z) |
180.0 → 110.1 |