Table 2.

Random and systematic errors coming from the practical evaluation of radiation

Source One standard deviation Δi
Measurement:
TLD counting statistics 3% 10%
Others:
WT (weighting factor) 5%
VMAT output photon dose < 2%
TLD-100H calibration 3% 20%
Inserting positions of phantom organ or tissue 5%
Density effects of PMMA phantom 1.5% 3.8%
Variation of Harshaw 3500 reader 3% 8%
Δtot 9.00% 21.0%