Skip to main content
. 2022 Apr 13;14(8):1626. doi: 10.3390/nu14081626

Table 2.

Selective validation parameters of the applied method.

Parameter Cd Pb Hg Ni
Reference value
(mg/kg)
0.30 ± 0.02 0.50 ± 0.04 0.20 ± 0.01 0.50 ± 0.03
Determined value (mg/kg) 0.27 0.53 0.16 0.52
0.31 0.48 0.18 0.51
0.33 0.46 0.16 0.57
0.29 0.52 0.21 0.49
0.27 0.52 0.23 0.55
0.28 0.54 0.17 0.5
Average 0.29 0.51 0.19 0.54
SD 0.02 0.03 0.03 0.03
RSD (%) 8.28 6.08 13.68 5.58
Recovery (%) 96.67 102.0 95.0 106.0
LOD * (µg/kg) 0.8 4.5 1.8 1.2
LOQ ** (µg/kg) 2.1 14.8 6.0 4.0

* LOD—Limit of Detection; ** LOQ—Limit of Quantification.