Ultra-high dose rate electron beams and the FLASH effect: from preclinical evidence to a new radiotherapy paradigm

Emil Schüler 1,2, Munjal Acharya 3, Pierre Montay-Gruel 4, Billy W Loo Jr 5, Marie-Catherine Vozenin 6,*, Peter G Maxim 4,*
PMCID: PMC9032195  NIHMSID: NIHMS1795857  PMID: 34997969

Abstract

In their seminal paper from 2014, Fauvadon et al. coined the term FLASH irradiation to describe ultra-high-dose-rate irradiation with dose rates greater than 40 Gy/s, which results in delivery times of fractions of a second. The experiments presented in that paper were performed with a high-dose-per-pulse 4.5-MeV electron beam, and the results served as the basis for the modern-day field of FLASH radiation therapy (RT). In this article, we review the studies that have been published after those early experiments, demonstrating the robust effects of FLASH RT on normal tissue sparing in preclinical models. We also outline the various irradiation parameters that have been used. Although the robustness of the biological response has been established, the mechanisms behind the FLASH effect are currently under investigation in a number of laboratories. However, differences in the magnitude of the FLASH effect between experiments in different labs have been reported. Reasons for these differences even within the same animal model are currently unknown, but likely has to do with the marked differences in irradiation parameter settings used. Here we show that these parameters are often not reported, which complicates large multi-study comparisons. For this reason, we propose a new standard for beam parameter reporting and discuss a systematic path to the clinical translation of FLASH radiation therapy.

1. FLASH Terminology and Critical Beam Parameters

The early definition of FLASH irradiation (mean dose rates (D˙¯) ≥ 40 Gy/s) has served as a good starting point for studying the dose rates at which the FLASH effect (normal tissue sparing with isoeffective tumor control after FLASH compared to conventional (CONV) irradiation) could be expected.1 Although FLASH irradiation has been shown to evoke strong, reproducible responses across many different organ systems (e.g., brain, lungs, gastrointestinal [GI] tract, skin) across multiple species2, variations in the magnitude of the FLASH effect has been reported between studies, and others have shown that ultra-high dose rate irradiation has either no or detrimental effects compared to CONV irradiation.39 Although these studies involved different types of radiation (X-rays, protons, and electrons), the common denominator was low mean dose rate, and 2 of the 3 studies used D˙¯ ≤ 40 Gy/s. Whether this could explain the lack of an observed FLASH effect is unknown at this time. However, understanding the intrinsic physical differences that exist between these positive and negative experiments should provide a better understanding of the FLASH effect.

A likely factor contributing to the variations in the magnitude of the FALSH effect between different studies could stem from inconsistencies in the physical radiation beam and fractionation parameters. Indeed, more recent findings suggest that the definition of D˙¯ > 40 Gy/s to achieve the FLASH effect is overly simplified; that the FLASH effect may also depend on other aspects of the radiation beam, potentially related to the total dose delivered, radiation source (e.g. electrons, protons, heavy particles, photons), irradiated volume, overall delivery time, and pulse-related factors such as dose and dose rate per pulse, as well as the frequency of the pulse delivery.10

To date, the most detailed characterizations of how the FLASH effect depends on the physical aspects of the radiation beam have been completed with mouse models of whole-brain irradiation. In one such series of experiments, the absorbed dose, pulse duration, and pulse frequency were kept constant (10 Gy, 1.8 μs, and 100 Hz, respectively), and the dose per pulse—and thus the mean dose rate, instantaneous dose rate, and irradiation duration—were varied.11 A sigmoidal response in neurocognitive performance (the endpoint in this model) was noted in which a dose-rate–dependent increase in memory was seen at D˙¯ > 18.5 Gy/s (instantaneous dose rate (D˙p): 1.0E5 Gy/s), with a plateau reached after D˙¯: 100 Gy/s (D˙p: 5.6E5 Gy/s). In another study of mouse whole brain irradiation, FLASH (D˙¯: 200–300 Gy/s, dose per pulse (Dp): 1.75 Gy, D˙p: 8.75E5 Gy/s) irradiation was found to produce less toxicity than CONV irradiation even when delivered with the same dose per pulse and number of pulses as FLASH (Dp: 1.75 Gy, 18 pulses) but over a longer total time (0.1 vs 240 seconds), suggesting the importance of overall delivery time.12 In 2019, a meta-analysis of available data on FLASH radiotherapy (RT) from in vivo models revealed the importance of overall irradiation time and dose rate within each pulse;13 this work was updated in 2021.14 How other physical characteristics of the radiation beam could affect the robustness and optimization of the FLASH effect remain unknown.

Whereas most of the data investigating the FLASH effect to date have been obtained with pulsed electron beams, robust comparisons among studies require certain parameters to be defined,15 such as beam energy, beam structure, total dose, mean dose rate, instantaneous (intra-pulse) dose rate, pulse repetition frequency, dose per pulse, pulse width, duration of exposure, field size, percentage depth dose, dose profiles, and irradiated volume (summarized in Table 1). Also crucial are specific definitions of where these parameters are defined, and what the dose gradient is across the irradiated volume of interest. For example, in mouse irradiations, the dose gradient for electron irradiations can be >15%/cm depending on the tissue type traversed by the beam, beam energy, field size, and the source-to-surface (SSD) distance.16 For this reason, it is important to have a common definition of how and where the dose and dose parameters are defined, and that the dosimetry is performed in a geometry that is representative of the experimental setup.

Table 1.

Reportable physical parameters of the electron FLASH beam

Irradiation parameters Unit of Measure Comments
Beam energy MeV
Total absorbed dose Gy Defined to a Dose Specification Point (DSP) at the center of the irradiated volume of interest. The coordinates of this point in relation to the geometry of interest must also be reported.
Fractionation schedule Dose per fraction, number of fractions, and time between fractions
Mean dose rate per fraction (D˙¯) Gy/s
Instantaneous dose rate (D˙p) Gy/s Dose rate within one pulse. Also reported is the variation in instantaneous dose rates within the pulse train delivered.
Pulse frequency Hz
Dose per pulse (Dp) Gy Mean dose per pulse and dose per pulse variation within the pulse train delivered
Pulse width s At Full Width at Half Maximum (FWHM)
Duration of exposure s Total irradiation time = number of pulses * pulse width + (number of pulses - 1) * pulse separation
Beam field size Field shape and area irradiated, defined at the surface of the irradiated geometry. E.g., AxB [mm2] for square fields, or diameter [mm] of circular fields.
Dose profiles % Dose profiles taken along the lateral and axial directions in the coordinate system defined and described by the researcher. The profiles are centered on and normalized to the DSP. Alternatively, two point-doses along the central axis should be reported (DSP + surface dose) in case no dose profiles are supplied.
Irradiated volume mm3 Volume receiving 10% or more of prescription dose

To date, different studies have used different definitions and approaches for dose determination. For a better understanding of the irradiation setup, and to facilitate for a broader comparison of the irradiation parameters used both temporally and spatially while still considering the limitations in hardware and software available, we suggest the following set to be at a minimum reported in terms of dose parameter: The dose and dose parameters should be defined to a Dose Specification Point (DSP) at the center of the irradiated volume of interest. If a highly irregular volume is considered, then a representative DSP in this volume should be defined and used. The reporting of the dose parameters should be accompanied by the coordinates of the DSP as well as dose profile measurements along the lateral and axial directions, centered on the DSP. These dose profiles should be taken in a geometry that closely resembles the experimental setup if determined experimentally. If it is not feasible to obtain the dose profiles, then at least one more dose point, in addition to the central point, should be defined along the central axis of each beam used to facilitate evaluations of the dose gradient across the volume of interest. In the published literature, the entrance dose has been used extensively to define the delivered dose. However, the relationship between surface dose and dose at depth will depend on the type of radiation, the energy, and the SSD. Nevertheless, in the absence of a full percentage depth dose, it is recommended that the entrance dose also be defined along with the dose at the center of the irradiated volume of interest.

For multi-beam deliveries, the parameters in Table 1 should be reported for each individual beam and as a composite for each fraction. In addition, the time between each beam delivery needs to be reported to allow determination of the overall treatment time for the entire fraction.

Carefully documented and clearly defined experimental conditions are essential for ensuring reproducibility and hence results that can be compared among studies of FLASH RT, which in turn is needed for the safe and reliable transfer of FLASH RT strategies into the clinic.15 This detailed reporting of the irradiation parameters is needed to elucidate the contributions of the physical beam parameters to the FLASH effect, because pooling of data from different studies will be required. Although many of the parameters presented in Table 1 can be derived from one another, explicit definitions of each will minimize downstream errors. Notably, these parameters reflect only the physical parameters of the beam; biological parameters, described in the following sections, are equally important.

1.1. Measuring physical beam parameters

Measuring many of the physical parameters in Table 1 would require the implementation of detectors and dosimetry protocols that are not routinely used at present. Many of the characteristics of an ideal radiation detector are, however, shared by both CONV- and FLASH radiation beams, and include tissue equivalence, energy independence, and non-perturbing qualities. Other characteristics of increasing importance in FLASH beams are dose-rate independence, in terms of both mean and instantaneous dose rate, and the temporal resolution of the detector.2,13 The dose per pulse and instantaneous dose rate for electron beams can be on the order of 20 Gy and 1012 Gy/s, respectively, which pose a challenge when conventional detectors and dosimetry protocols are used due to e.g. saturation effects and loss of signal due to the rapid energy deposition in the detectors and dosimeters.17,18 Related to the temporal resolution of the detector, the ability to resolve individual pulses is needed to allow dose monitoring and beam control. In addition to the short time scale of resolving individual pulses, real-time (as opposed to passive) dose monitoring also becomes increasingly important in the development of the new control systems needed to translate FLASH-capable machines into clinical use.

In relation to dose-rate–dependence is also the dynamic range of the detector. Ideally, the dynamic range should cover both the lower and upper limits of both CONV- and FLASH irradiation to facilitate the use of a single detector for both types of irradiation without having to resort to excessive corrections to the readings. Spatial resolution is also vital, particularly in preclinical models that involve the use of extremely small fields; such irradiations fall into the realm of “small-field dosimetry” owing to the loss of the lateral charged particle equilibrium.19 Volume averaging becomes a concern in small fields, as larger detectors would be exposed to different dose rates across different parts of the detector volume.

An in-depth review of the available detector systems and their applicability in FLASH dosimetry is beyond the scope of this review; interested readers are referred to another review by Ashraf et al.18 However, certain detector systems are worth mentioning because of their universal use in FLASH dosimetry. Gafchromic film is extensively used as a dosimetry system in FLASH-related studies and due to its dose-rate independence is often used as the reference which to compare other detector systems to.20,21 Gafchromic film is preferred for its flexibility, tissue equivalency, energy independence (>50–100 kV), and high spatial resolution. Gafchromic film has been extensively characterized in high dose rate beams and has been found to be independent of dose rates up to 9E12 Gy/s.20,21 However, Gafchromic film’s major drawback is that it is limited by delays in the readout; ideally the film should not be analyzed until 24 h after the irradiation to allow the rate of polymerization to stabilize.18 Other passive dosimeters heavily used in FLASH dosimetry include alanine and thermoluminescent dosimeters (TLDs).22,23 These detectors share many of the characteristics of Gafchromic film but are limited to point-dose measurements. Nevertheless, alanine and TLDs have been shown to be independent of dose rates of >1.5E9 Gy/s and are very useful for the characterization of FLASH beams in in vivo dosimetry.20,24

Ion chambers, which are an integral part of normal clinical operations in CONV RT, are challenging to use with FLASH beams. As a large amount of energy is transferred within a single pulse to the collection volume of the chamber, the collection efficiency of ion chambers is reduced by recombination of the generated ion pairs before being collected by the electrodes. Models that are typically used to account for recombination are not applicable to FLASH dose rates.18 However, Petersson et al.25 attempted to circumvent this limitation. With their detailed characterization of the Advanced Marcus chamber and with use of Gafchromic film as the standard, this group developed a logistic function for ion recombination correction that can be used successfully in FLASH beam lines.

Perhaps the most promising detectors for use in FLASH beam dosimetry are scintillator- and Cherenkov-based detectors. Organic scintillator detectors are tissue-equivalent, energy- and dose-rate–independent, and allow measurement of integrated dose in real time as well as temporal resolution of individual linac pulses.2628 Inorganic scintillator detectors have similar characteristics, but their use of high atomic number materials preclude their tissue equivalence and energy independence. However, because these detectors are generally more radiation-resistant, and have higher light output and easier corrections for the stem-effect than organic scintillators, they remain attractive for FLASH beam dosimetry. For FLASH dosimetry, the signal generated would be several orders of magnitude higher than that generated in CONV dose-rate dosimetry. For this reason, the active volume could be made smaller and allow easier filtering of signal pertaining to stem-effect features, even with organic scintillators. Archer et al. recently demonstrated the successful implementation of a 10-μm thick BC-400 film plastic scintillator coupled to a 1-mm diameter optical fiber in a synchrotron X-ray beam, which showed excellent performance at dose rates of up to 4435 Gy/s.29 Two- and three-dimensional measurements with luminescent technology have also been successfully implemented.3034

2. FLASH: Preclinical Investigations using Electron Beams

2.1. Normal tissues

The FLASH effect, as noted above, is defined by the preservation of normal tissues simultaneous with antitumor efficacy equivalent to that of CONV RT at the same dose level.1,12,13,15,22,23,3539 To date, the FLASH effect has been characterized in several in vivo models, primarily wild-type mice, and in several organ systems, as summarized in Table 2. These organs include either the so-called acute-responding organs (gut, hematopoietic system)36,40 as well as late-responding organs (brain, lung, skin),1,12,13,15,22,23,35,37,38,41,42 thereby suggesting that FLASH RT modifies a common initial event that can control the development of both acute and delayed toxicity. The fundamental physicochemical mechanisms underlying the FLASH effect are currently under investigation; one hypothesis implicating transient local oxygen depletion was first proposed nearly 40 years ago.43,44 However, direct measurements of such depletion in vivo are difficult to obtain,45,46 and so most of the evidence to date has been collected via indirect measurements. As one example, reductions in hydrogen peroxide (H2O2) in FLASH irradiated water have been reported.35 This result, considered with the lack of radioprotection triggered by antioxidants in FLASH irradiated zebrafish embryos,13,35 suggests that FLASH-RT can modify the initial rate or production of ROS in tissues, which in turn can affect the entire downstream biological cascade. Some important biological mechanisms, downstream to the ROS-mediated events, known to be involved in the FLASH effect on normal tissues include reduction of DNA damage (lung),41 apoptosis (lung and brain),1,12 and pro-inflammatory consequences (brain)12,47 (Table 2).

Table 2.

In vivo studies of electron beams to investigate FLASH effects in normal and cancerous tissues (in order of publication year)

In vivo studies Irradiation parameters
Reference Model / Species Target site or organ Biological endpoint Total dose (Gy) Beam energy (MeV) Delivery time (s) Mean dose rate, D˙- (Gy/s) Instantaneous dose rate, Dp˙ (Gy/s) Number of pulses Pulse rate (Hz)
Field 19741 Rat Skin and hind feet Early skin reactions; Late deformities 20 – 35 7 2.5 × 10−1 – 5 × 10−1 67 – 80 Not specified Not specified Not specified
Inada 19802 Mouse (ICR) Skin (Right hind leg) Early skin reactions 30, 50 8 12 – 20 2.5 >1.6 × 104 Not specified 23
Hendry 19823 Mouse (C57Bl/6J) Tail skin Skin necrosis (ND50); skin regeneration; epithelial integrity 30, 50 10 1.8 × 10−1 – 2.9 17 – 170 1 × 105 – 1 × 106 Not specified 50
Favaudon 20144 Mouse (C57Bl/6J; Swiss Nude) Lung Lung fibrosis, tumor kill, growth delay 17 – 30 4.5 > 1 × 10−1 40 – 60 1 × 106 4 – 6 100 – 150
Montay-Gruel 20175 Mouse (C57Bl/6J) Whole brain and hippocampi Behavior (NOR); cell proliferation (BrdU) 10 6 1.8 × 10−6 – 0.3 33 – 5.6 × 106 1.9 × 105 – 5.6 × 106 1 – 10 100
Vozenin 20196 Zebrafish embryo Embryos Fish length 5 – 12 6 1.8 × 10−6 > 2.7 × 106 > 4.4 × 106 1 Single pulse
Montay-Gruel 20197 Zebrafish embryo Embryos Fish length 8 6 1.8 × 10−6 4.4 × 106 4.4 × 106 1 Single pulse
Vozenin 20198 Mini pig Skin (dorsal) Fibronecrotic lesions 28, 31, 34 6 0,1 > 280 1.8 × 106 10 100
Vozenin 20198 Cat* Nose Squamous cell carcinoma Skin macroscopic complete response 25 – 41 4.5 and 6 > 9.0 × 10−2 > 277 0.5 × 106 – 1.8 × 106 9–10 100
Simmons 20199 Mouse (C57Bl/6J) Whole brain Dendritic spines (hippocampal neurons), neuroinflammation, brain cytokines 30 16 – 20 1.0 × 10−1 – 1.6 × 10−1 200, 300 8.75 × 105 18 106, 180
Montay-Gruel 20197 Mouse (C57Bl/6J) Whole brain Cognition, anxiety, neuron structure, neuroinflammation, gliosis 10 – 14 6 1.8 × 10−6 5.6 × 106 – 7.8 × 106 5.6 × 106 – 7.8 × 106 1 Single pulse
Bourhis 201910 Human Patient CD30+ T-cell cutaneous lymphoma Cancerous and normal skin Tumor response, normal skin health 15 6 9.0 × 10−2 158 1.5 × 106 10 100
Venkatesulu 2019, 202011, 12 Mouse (C57Bl/6J, BALB/C) Thoracic, splenic, abdominal Lymphopenia parameters, survival 10 – 16 20 2.6 × 10−1 – 4.9 × 10−1 32.6, 38.8 Not specified Not specified 180
Fouillade 202013 Mouse (C57BL/6J, Terc−/−) Lung DNA damage markers, cell proliferation, senescence, RNA-seq (inflammation genes) 17 4.5 3.3 × 10−2 – 1.1 × 10−1 135 – 600 8 × 105 – 3.2 × 106 5 – 11 100 – 150
Chabi 202014 Mouse NSG (PDX T-ALL model) Total body irradiation Patient T-ALL IV injection Hematopoiesis, survival, gene expression array (cancer markers) 4 6 2 × 10−2 200 7.4 ×105 3 100
Montay-Gruel 202015 Mouse (C57Bl/6J; Swiss Nude) Whole brain and hemi brain Tumor (GBM) growth, survival, behavior (NOR) 10, 14, 25, 2 × 7, 4 × 3.5 6 1.0 × 10−2 – 1.8 × 10−6 2.5 × 103 – 7.8 × 106 1.9 × 106 – 7.8 × 106 1 – 2 100
Alaghband 202016 Juvenile mouse (C57Bl/6J) Whole brain Cognition, neurogenesis, neuroinflammation 8 6 1.8 × 10−6 4.4 × 106 4.4 × 106 1 Single pulse
Montay-Gruel 202017 Mouse (C57Bl/6J) Whole brain Astrogliosis, complement activation, TLR4 10 6 1.8 × 10−6 5.6 × 106 5.6 × 106 1 Single pulse
Allen 202018 Mouse (C57Bl/6J) Whole brain Brain microvasculature 10, 25 6 10−2 – 1.8 × 10−6 2.5 × 103 – 1.8 × 106 5.6 × 106 – 6.9 × 106 1 – 2 100
Kim 202019 Mouse (C57Bl/6J) Subcutaneous Lewis lung carcinoma Tumor vascular morphology, ROS, immune cell infiltration, DNA damage 14 16 3.9 × 10−2 352 4.3 × 105 8 100
Soto 202020 Mouse (C57Bl/6J) Hemi-thorax Skin health, survival 10 – 40 16 5.5 × 10−2 – 2.2 × 10−1 180 4.0 × 105 5 – 20 90
Levy 202021 Mouse (C57Bl/6J) Abdominal GI syndrome, survival, crypt cell regeneration, epithelial integrity, tumor growth 14 – 16 16 6.5 × 10−2 – 7.4 × 10−2 216 4.0 × 105 7 – 8 108
*

Veterinary patient study

2.2. Tumors

To date, data derived from single-beam, single-dose studies provide convincing evidence that FLASH RT is isoefficient relative to CONV irradiation at controlling tumor growth rates. This observation has been reported in various xenograft models (breast, prostate, lung, glioblastoma [GBM])1 and in orthotopic tumor models (lung, GBM)13,36,48 as well as in transgenic mice.36,48 More recently, patient-derived xenograft (PDX) models of T-cell acute lymphoblastic leukemia have shown tumor-subtype–specific susceptibility to FLASH RT.40 Because fractionated RT regimens are the standard of care for the treatment of solid tumors, it is also important to determine whether dose fractionation would preserve the FLASH effect. To address this directly, several hypo-fractionated regimens (3.5 Gy x 4, 7 Gy x 2, and 10 Gy x 3) were found to produce no distinguishable differences in overall survival and tumor growth delay after FLASH- vs CONV irradiation.14 Notably, large single-dose regimens (a single 10 Gy fraction) and hypo-fractionated regimens (doses of >=7 Gy per fraction) could still spare normal tissues without affecting antitumor effectiveness.

The difference in the reactions of normal tissues vs tumors to FLASH RT provides a unique opportunity for the field of radiation oncology to enhance tumor control safely and more effectively. Moreover, as radiation oncology shifts to using more hypo-fractionated regimens to treat a variety of tumor types, the capability to dose-escalate with FLASH RT facilitates these approaches and minimizes the number of patient visits, which will ultimately reduce the cost of healthcare.

3. High-throughput model systems for investigating beam parameters crucial for achieving the maximum FLASH effect

The ideal biological system for investigating how physical beam parameters influence the FLASH effect would have fast readout, high throughput, and robust and reproducible responses. The standard approach to fulfilling these requirements has historically been through detailed in vitro studies, primarily clonogenic assays.49 However, because the FLASH effect is defined in terms of sparing of normal tissues after a dose delivered at ultra-high dose rate relative to CONV dose rates, then by definition, the FLASH effect should be evaluated in vivo. Indeed, testing proposed hypotheses regarding the mechanisms of FLASH, with the likelihood that the FLASH effect is a combination of several effects, makes it crucial that beam validation is done in vivo before undertaking any mechanistic studies in vitro. However, in vivo studies have some major drawbacks relative to in vitro studies including general animal experimentation ethics and handling issues, increased complexity of the response, and longer readout time for the results. The realization that the FLASH effect is related to specific irradiation parameters, and that different organ systems are likely to require their own optimized set of parameters, underscores the need to find robust, high-throughput and reproducible in vivo systems to properly evaluate how the numerous possible combinations of beam parameters can maximize the FLASH effect.

The potential of high-throughput screening of the FLASH effect in an in vivo system is perhaps best illustrated with the zebrafish model, in which embryos are exposed to radiation after fertilization and then assessed for viability and morphologic abnormalities. The benefits of this model include ease of handling, rapid development, and ease of visualizing major organs.50 The relevance and responsiveness of early-stage zebrafish embryos (4 hours after fertilization) to ultra-high dose rate irradiation was validated by using an eRT6 electron beam51 (Figure 1) and subsequently used to evaluate FLASH RT in combination with the reactive oxygen species (ROS) scavenger amifostine (single pulse delivery, D˙¯=D˙p:5.67.8E6Gy/s, Dp: 10–14 Gy).35 On the other hand, Beyreuther et al.4 failed to observe any difference in survival or morphologic integrity when zebrafish embryos at a later developmental stage (24 hours after fertilization) were exposed to a proton beam line at the Proton Therapy University of Dresden. Only reductions in pericardial edema after the delivery of 23 Gy with FLASH (D˙¯: 100 Gy/s, D˙p: 0.5E3 Gy/s) compared to CONV irradiation (D˙¯: 0.08 Gy/s, D˙p: 0.4 Gy/s), were reported.4 More recently, the same group using the same model succeeded in protecting zebrafish embryo development by using a research electron beam line (Electron Beam of high Brilliance and low Emittance [ELBE]52) operating at a higher dose rate (CONV: D˙¯: 0.11 Gy/s (continuous delivery); FLASH: D˙¯: 10E5 Gy/s, Dp: 1.8E-2 Gy D˙p: 10E9 Gy/s).53 The significance of these discrepant results is under investigation but may be related to the mean and instantaneous dose rates used and the nature of the beam.

Figure 1: Zebrafish embryos irradiated with the eRT6 electron beam FLASH did not show developmental retardation.

Figure 1:

At 4 hours after fertilization, zebrafish embryos were irradiated and their development was analyzed 5 days later. Below: Normal development of zebrafish embryos (a) was altered following CONV irradiation (b) but not after FLASH irradiation (c) Right panel: dose response curve after CONV and FLASH-RT. Mean standard deviation and Mann-Whitney’s-test (at 10 Gy P < 0.01; at 11–12 Gy P < 0.001 (n =20 embryos/group).

Another system with the capability of high-throughput readout that has been extensively described in the literature is the Withers-Elkind crypt assay, also known as the microcolony assay. This assay, first described in 1970,54 monitors the regeneration of intestinal crypts after irradiation. In brief, mice are killed 3.5 days after irradiation of the GU tract; segments of the jejunum are removed and processed by routine histologic techniques; and transverse sections are cut and stained with hematoxylin and eosin. The numbers of regenerating crypts are counted and presented as regenerating crypts per circumference. Regenerative crypts are generally scored visually; Withers and Elkind defined objective criteria for regenerative crypts as “10 or more cells, each with a prominent nucleus and little cytoplasm, lying close together and appearing crowded”. Levy et al. used this assay to study the different effects of total abdominal irradiation given by FLASH vs CONV irradiation.36 FLASH irradiation (D˙¯: 216 Gy/s, Dp: 2 Gy, D˙p: 4E5 Gy/s) induced significantly less intestinal injury (increased survival and decreased death of crypt base columnar cells) in healthy mice compared to CONV irradiated mice. Moreover, FLASH irradiation increased sparing of normal tissue while retaining tumor control after total abdominal irradiation in the ovarian cancer model ID8 compared to CONV irradiated mice.

Notably, however, zebrafish embryos and the microcolony assay both involve use of highly proliferative cells that model the response of the acute responding organ well, but they are not relevant for investigating delayed response, suggesting that the best model with which to validate the FLASH effect at this time is still mice.

Although cell culture has been used widely for radiobiology studies over the years, in vitro experimentation to validate the FLASH effect does not substitute for in vivo functional validation. For instance, past43,44 and present studies55 have shown that in vitro, radioprotection was not observed under atmospheric conditions (21% O2) but was observed in hypoxic conditions at doses around 20 Gy. Given the very low surviving fraction obtained at such high doses (< 0.001 in hypoxic environment)55, the relevance of these in vitro studies with such high doses for observing the FLASH effect is unclear. More globally, the relevance of 2D cultures to investigate radiation response to FLASH-RT can be questioned for many reasons. For example, use of FLASH can induce radioprotection in tumor cell lines irradiated in vitro, whereas FLASH RT does not protect tumors in vivo1,36,48(see section 2.2). Under standard clonogenic conditions (doses of 2–6 Gy and atmospheric dioxygen tension), no differential effects have been noted thus far between normal and tumor cells exposed to FLASH RT in vitro (Figure 2).

Figure 2. Clonogenic assays after irradiation with the eRT6 electron beam.

Figure 2.

Three tumor cell lines (human GBM U87, human cervix HeLa, murine GBM H454) and one normal cell line (HaCat) were irradiated with FLASH and CONV RT. No differences in clonogenic survival were measured between FLASH and CONV irradiation in cancer cell lines and one normal cell line (HaCat).

Recently, 3D culture models such as spheroids, organoids, or even more complex organ-on-a-chip systems have emerged in radiobiology, and their relevance for studying the FLASH effect has yet to be explored.42 Some potential advantages are obvious, such as the opportunity to investigate the effects of FLASH RT on human cells rather than rodent cells. The microenvironment in such models is also complex, and the paracrine signals between the various cellular compartments are maintained (e.g., crosstalk between vasculature and cell types as well as circulating blood and immune cells). Other advantages of such models are the ease of modeling several physiobiological equations, such as the diffusion and metabolism of oxygen within the spheroids; its depletion through reactions involving radiation-induced radicals; and the increase in radio-resistance using the classical models of oxygen enhancement ratio and linear-quadratic response. Subsequently, these models can be readily verified by evaluating growth after irradiation. While spheroids provide an innovative model to study FLASH radiation therapy, both computationally and experimentally, it should be noted that the relevance to in vivo tumors is unclear, thus results from such studies need to be evaluated carefully and placed into context.

4. Path to Clinical Translation: Models Relevant to Clinical Questions

Scaling FLASH technology from conditions and geometries that work for rodents to those that would be applicable to human patients is a substantial challenge. For example, exploring the effects of volume and conformality would require experiments with large animals. To this end, pigs have been used in radiopathology and radio-oncology for decades,6365,66 and a previous study have shown that pigs are suitable for investigating normal tissue responses to FLASH-RT as well as being useful for comparing normal skin response to CONV- vs FLASH RT.23 However, investigations of tumor response ideally require model systems in which cancers arise spontaneously. Such model systems may already be available through veterinary practice. Indeed, RT has become an essential part of cancer treatment in animals23,67; the advances in high-precision treatment delivery and multimodal imaging used for human patients are increasingly being translated to veterinary practice, thereby providing opportunities to use domesticated animals with spontaneously arising tumors to explore the potential benefits of FLASH RT.23,67 Radiation effects have been well studied in cats and dogs6870; these animals could serve as models for testing the safety of FLASH RT.23 As one example, a phase III randomized trial of FLASH vs CONV RT for cats with squamous cell carcinoma of the nasal planum is currently ongoing at the CHUV (Lausanne University Hospital) in collaboration with Prof Dr. C Rohrer-Bley at the University of Zurich. The aim of this trial is to compare tumor control and short-term and long-term toxicity between standard-of-care CONV RT (4.8 Gy x 10) and a single 30-Gy dose of FLASH-RT. Other feasibility experiments in dogs with osteosarcoma and sarcoma are underway, one with proton-FLASH at the University of Pennsylvania and another at the University of Lund with electron-FLASH.71 Unfortunately, these trials do not include a control condition in which RT is given at CONV dose rates.

In summary, using FLASH RT to treat companion animals with cancer may be important not only as an intermediate step toward applying this technology to human patients but also could be beneficial for the animals as well. For example, the ability to deliver FLASH RT in hypo-fractionated regimens—and even in single fractions—would be expected to enhance the pet patient’s quality of life during treatment (because of the need for fewer anesthesia sessions) and after treatment (tumor control without normal-tissue toxicity). FLASH RT may also make RT more affordable and reduce the workload at veterinary clinics. Finally, trials of FLASH RT in companion animals can also be useful for designing clinical workflows for future studies with patients.

5. Technological Perspectives on Future Developments

Cancer is predicted to be the leading cause of death worldwide, with about 30.2 million newly diagnosed cases and 16.3 million related deaths per year by the year 2040.72 These numbers underscore the need for innovative treatment modalities against cancer and warrant ways of meeting the associated major challenges for 21st century health care. In that context, the tantalizing possibility that the FLASH effect crosses tissues and species, and the magnitude of the benefit observed in various preclinical studies, highlights the need to define and promote its clinical application.2,10,13,73,74 Typically, the speed at which FLASH-RT is delivered may be sufficient to circumvent problems with organ motion during treatment in the setting of real-time imaging, which is otherwise an important consideration in CONV RT. However, the required parameters for clinical application need to be carefully defined and tested in proof-of-safety trials before FLASH RT can be widely applied in clinical settings.

From a technological point of view, only a few systems exist at present that can operate at the ultra-high dose rates associated with the FLASH effect. Although most FLASH studies to date have been performed with experimental 4- to 6-MeV electron devices13,15 or modified clinical linacs,16,75,76 the use of synchrotron X-rays for FLASH-RT has also been studied,77 as has the use of proton beams.4,78,79 In 2021, the CHUV/Lausanne University Hospital team plans to begin using two electron beams of about 9 MeV (Mobetron/IntraOp and FLASHknife/PMB-Alcen) to treat superficial skin cancers and for intraoperative treatment of other types of cancer.80 This pragmatic approach is expected to provide the proof of principle as to a FLASH RT benefit in humans before the development of devices for treating deep-seated tumors. Another approach being developed as a collaboration between CHUV and CERN, the European Council for Nuclear Research in Geneva (https://cerncourier.com/a/adapting-clic-tech-for-flash-therapy), aims to produce a very high energy electron (VHEE) - FLASH device that can deliver high doses at high dose rates to relatively large volumes and deep-seated tumors.8184 VHEE beams offer both the penetration needed and the penumbra that is practical for deep seated tumors.81,85,86 VHEE beams also have the advantage of increased uniform dose at high density boundaries compared to photon beams, and compared to protons, allow for easy electromagnetic scanning. The potential disadvantages of using VHEE beams depends on the proposed technique, but with the currently proposed systems there is added complexity and cost compared to conventional clinical technology, which will limit the widespread availability of this technology in the near future.

Conclusion

The field of FLASH RT is still in its infancy and the true potential of this novel treatment strategy is still to be determined. In order for us to determine the true potential of FLASH RT we need to understand what constitutes FLASH irradiation in terms of the physical beam parameters needed to induce the FLASH effect. We also need to understand what effects the manipulation of the physical beam parameters have on the magnitude of the FLASH effect. This knowledge can then be used to elucidate the underlying biological mechanism(s) of the FLASH effect and allow for an optimization of FLASH RT in terms of normal tissue sparing to critical organ systems. The first step to get to this point is for the community to agree upon a common reporting system of the critical beam parameters used to allow for transparency and retrospective studies of all aspects of the beam delivery and experimental setup. In this review, we are proposing a set of parameters to be reported in future studies within the field of FLASH RT. This list constitutes the first step to a common reporting system and can be easily adapted also to other irradiation types beyond electron irradiations. This reporting system will likely need to be modified as the field moves forward, and more advanced delivery systems are developed. However, if adopted, it would constitute the first step towards a new standard of beam parameter reporting and facilitate for the robust and safe translation of this technology to the clinical setting.

Acknowledgement:

The authors would like to thank Christine F. Wogan, MS, ELS, of MD Anderson’s Division of Radiation Oncology, for editorial contributions to several drafts of this article. The authors would also like to thank the FLASH team of the CHUV: B. Petit and J. Ollivier were involved in the biology studies; C. Bailat, F. Bochud, J-F Germond, R. Moeckli, P. Goncalves Jorge in the physics studies as well as J Bourhis. We would like to thank the animal facilities of Epalinges and zebrafish platform of CHUV/UNIL. The FLASH studies are supported by grants from the Swiss National fund SNF Synergia FNS CRS II5_186369 (MCV) and NIH program project grant PO1CA244091 (MCV, BL, PM). MMA was supported by American Cancer Society (ACS) Research Scholar grant (RSG-17-146-01-CCE) and UCI Institute for Clinical and Translational Sciences KL2 (KL2TR001416) awards. ES was supported by the University Cancer Foundation via the Institutional Research Grant program at the University of Texas MD Anderson Cancer Center, by a grant from The University of Texas MD Anderson Cancer Center, Division of Radiation Oncology, and in part by Cancer Center Support Grant P30 CA016672 from the National Cancer Institute of the National Institutes of Health, to The University of Texas MD Anderson Cancer Center.

Disclosures

BWL has received research support from Varian Medical Systems; he is a co-founder and board member of TibaRay. ES has received research support from IntraOp Medical. MCV has received research support from Varian Medical Systems in the frame of a collaborative work with Paul Scherrer Institute (PSI), she was member of Scientific Advisory Board for Biowin/IBA project and is receiving research support from ROCHE pharma.

REFERENCES

  • 1.Favaudon V, Caplier L, Monceau V, et al. Ultrahigh dose-rate FLASH irradiation increases the differential response between normal and tumor tissue in mice. Science translational medicine. 2014;6(245):245ra293–245ra293. [DOI] [PubMed] [Google Scholar]
  • 2.Wilson JD, Hammond EM, Higgins GS, Petersson K. Ultra-high dose rate (FLASH) radiotherapy: Silver bullet or fool’s gold? Frontiers in Oncology. 2020;9:1563. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Smyth LM, Donoghue JF, Ventura JA, et al. Comparative toxicity of synchrotron and conventional radiation therapy based on total and partial body irradiation in a murine model. Scientific reports. 2018;8(1):1–11. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Beyreuther E, Brand M, Hans S, et al. Feasibility of proton FLASH effect tested by zebrafish embryo irradiation. Radiotherapy and Oncology. 2019;139:46–50. [DOI] [PubMed] [Google Scholar]
  • 5.Venkatesulu BP, Sharma A, Pollard-Larkin JM, et al. Ultra high dose rate (35 Gy/sec) radiation does not spare the normal tissue in cardiac and splenic models of lymphopenia and gastrointestinal syndrome. Scientific reports. 2019;9(1):1–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Beyreuther E, Brand M, Hans S, et al. Feasibility of proton FLASH effect tested by zebrafish embryo irradiation [published online ahead of print 2019/07/04]. Radiother Oncol. 2019;139:46–50. [DOI] [PubMed] [Google Scholar]
  • 7.Beyreuther E, Karsch L, Laschinsky L, et al. Radiobiological response to ultra-short pulsed megavoltage electron beams of ultra-high pulse dose rate [published online ahead of print 2015/05/15]. Int J Radiat Biol. 2015;91(8):643–652. [DOI] [PubMed] [Google Scholar]
  • 8.Oppelt M, Baumann M, Bergmann R, et al. Comparison study of in vivo dose response to laser-driven versus conventional electron beam [published online ahead of print 2015/01/21]. Radiat Environ Biophys. 2015;54(2):155–166. [DOI] [PubMed] [Google Scholar]
  • 9.Smyth LML, Donoghue JF, Ventura JA, et al. Comparative toxicity of synchrotron and conventional radiation therapy based on total and partial body irradiation in a murine model [published online ahead of print 2018/08/15]. Sci Rep. 2018;8(1):12044. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Vozenin M-C, Hendry JH, Limoli C. Biological benefits of ultra-high dose rate FLASH radiotherapy: sleeping beauty awoken. Clinical oncology. 2019;31(7):407–415. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Montay-Gruel P, Petersson K, Jaccard M, et al. Irradiation in a flash: Unique sparing of memory in mice after whole brain irradiation with dose rates above 100 Gy/s. Radiotherapy and Oncology. 2017;124(3):365–369. [DOI] [PubMed] [Google Scholar]
  • 12.Simmons DA, Lartey FM, Schüler E, et al. Reduced cognitive deficits after FLASH irradiation of whole mouse brain are associated with less hippocampal dendritic spine loss and neuroinflammation. Radiotherapy and Oncology. 2019;139:4–10. [DOI] [PubMed] [Google Scholar]
  • 13.Bourhis J, Montay-Gruel P, Jorge PG, et al. Clinical translation of FLASH radiotherapy: Why and how? Radiotherapy and Oncology. 2019;139:11–17. [DOI] [PubMed] [Google Scholar]
  • 14.Montay-Gruel P, Acharya MM, Jorge PG, et al. Hypofractionated FLASH-RT as an effective treatment against glioblastoma that reduces neurocognitive side effects in mice. Clinical Cancer Research. 2021;27(3):775–784. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Vozenin M-C, Montay-Gruel P, Limoli C, Germond J-F. All Irradiations that are Ultra-High Dose Rate may not be FLASH: The Critical Importance of Beam Parameter Characterization and In Vivo Validation of the FLASH Effect. Radiation Research. 2020. doi: 10.1667/rade-20-00141.1. [DOI] [PubMed] [Google Scholar]
  • 16.Schüler E, Trovati S, King G, et al. Experimental platform for ultra-high dose rate FLASH irradiation of small animals using a clinical linear accelerator. International Journal of Radiation Oncology* Biology* Physics. 2017;97(1):195–203. [DOI] [PubMed] [Google Scholar]
  • 17.Esplen NM, Mendonca MS, Bazalova-Carter M. Physics and biology of ultrahigh dose-rate (FLASH) radiotherapy: a topical review. Physics in Medicine & Biology. 2020. [DOI] [PubMed] [Google Scholar]
  • 18.Ashraf MR, Rahman M, Zhang R, et al. Dosimetry for FLASH Radiotherapy: A Review of Tools and the Role of Radioluminescence and Cherenkov Emission. Frontiers in Physics. 2020;8(328). [Google Scholar]
  • 19.IAEA TRS-483 IAEA. Dosimetry of Small Static Fields Used in External Beam Radiotherapy, Technical Report Series No. 483. Vienna: INTERNATIONAL ATOMIC ENERGY AGENCY; 2017. [Google Scholar]
  • 20.Karsch L, Beyreuther E, Burris‐Mog T, et al. Dose rate dependence for different dosimeters and detectors: TLD, OSL, EBT films, and diamond detectors. Medical physics. 2012;39(5):2447–2455. [DOI] [PubMed] [Google Scholar]
  • 21.Bazalova‐Carter M, Liu M, Palma B, et al. Comparison of film measurements and Monte Carlo simulations of dose delivered with very high‐energy electron beams in a polystyrene phantom. Medical physics. 2015;42(4):1606–1613. [DOI] [PubMed] [Google Scholar]
  • 22.Jorge PG, Jaccard M, Petersson K, et al. Dosimetric and preparation procedures for irradiating biological models with pulsed electron beam at ultra-high dose-rate. Radiotherapy and Oncology. 2019;139:34–39. [DOI] [PubMed] [Google Scholar]
  • 23.Vozenin M-C, De Fornel P, Petersson K, et al. The advantage of FLASH radiotherapy confirmed in mini-pig and cat-cancer patients. Clinical Cancer Research. 2019;25(1):35–42. [DOI] [PubMed] [Google Scholar]
  • 24.Kudoh H, Celina M, Kaye R, Gillen K, Clough R. Response of alanine dosimeters at very high dose rate. Applied radiation and isotopes. 1997;48(4):497–499. [Google Scholar]
  • 25.Petersson K, Jaccard M, Germond JF, et al. High dose‐per‐pulse electron beam dosimetry—A model to correct for the ion recombination in the Advanced Markus ionization chamber. Medical physics. 2017;44(3):1157–1167. [DOI] [PubMed] [Google Scholar]
  • 26.Beaulieu L, Beddar S. Review of plastic and liquid scintillation dosimetry for photon, electron, and proton therapy. Physics in Medicine & Biology. 2016;61(20):R305. [DOI] [PubMed] [Google Scholar]
  • 27.Beddar A, Mackie T, Attix F. Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical considerations. Physics in Medicine & Biology. 1992;37(10):1883. [DOI] [PubMed] [Google Scholar]
  • 28.Beddar AS, Mackie T, Attix F. Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurements. Physics in Medicine & Biology. 1992;37(10):1901. [DOI] [PubMed] [Google Scholar]
  • 29.Archer J, Li E, Davis J, Cameron M, Rosenfeld A, Lerch M. High spatial resolution scintillator dosimetry of synchrotron microbeams. Scientific reports. 2019;9(1):1–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Ashraf MR, Bruza P, Pogue BW, et al. Optical imaging provides rapid verification of static small beams, radiosurgery, and VMAT plans with millimeter resolution. Medical physics. 2019;46(11):5227–5237. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Vigdor S, Klyachko A, Solberg K, Pankuch M. A gas scintillator detector for 2D dose profile monitoring in pencil beam scanning and pulsed beam proton radiotherapy treatments. Physics in Medicine & Biology. 2017;62(12):4946. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Darne C, Alsanea F, Robertson D, Hojo Y, Sahoo N, Beddar S. 3D Scintillator Detector System for Proton Scanning Beam Therapy. International Journal of Radiation Oncology• Biology• Physics. 2019;105(1):S89–S90. [Google Scholar]
  • 33.Darne CD, Alsanea F, Robertson DG, Sahoo N, Beddar S. Performance characterization of a 3D liquid scintillation detector for discrete spot scanning proton beam systems. Physics in Medicine & Biology. 2017;62(14):5652. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Goulet M, Rilling M, Gingras L, Beddar S, Beaulieu L, Archambault L. Novel, full 3D scintillation dosimetry using a static plenoptic camera. Medical physics. 2014;41(8Part1):082101. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Montay-Gruel P, Acharya MM, Petersson K, et al. Long-term neurocognitive benefits of FLASH radiotherapy driven by reduced reactive oxygen species [published online ahead of print 2019/05/18]. Proc Natl Acad Sci U S A. 2019;116(22):10943–10951. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Levy K, Natarajan S, Wang J, et al. Abdominal FLASH irradiation reduces radiation-induced gastrointestinal toxicity for the treatment of ovarian cancer in mice. Scientific Reports. 2020;10(1):21600–21600. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Ko RB, Soto LA, von Eyben R, et al. Evaluating the Reproducibility of Mouse Anatomy under Rotation in a Custom Immobilization Device for Conformal FLASH Radiotherapy [published online ahead of print 2020/08/29]. Radiat Res. 2020. doi: 10.1667/RADE-20-00095. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Soto LA, Casey KM, Wang J, et al. FLASH Irradiation Results in Reduced Severe Skin Toxicity Compared to Conventional-Dose-Rate Irradiation. Radiation Research. 2020;194(6):618–624, 617. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Kim YE, Gwak SH, Hong BJ, et al. Effects of ultra-high dose rate FLASH irradiation on the tumor microenvironment in Lewis lung carcinoma: role of myosin light chain [published online ahead of print 2020/11/14]. Int J Radiat Oncol Biol Phys. 2020. doi: 10.1016/j.ijrobp.2020.11.012. [DOI] [PubMed] [Google Scholar]
  • 40.Chabi S, To THV, Leavitt R, et al. Ultra-high-dose-rate FLASH and Conventional-Dose-Rate Irradiation Differentially Affect Human Acute Lymphoblastic Leukemia and Normal Hematopoiesis [published online ahead of print 2020/10/20]. Int J Radiat Oncol Biol Phys. 2021;109(3):819–829. [DOI] [PubMed] [Google Scholar]
  • 41.Fouillade C, Curras-Alonso S, Giuranno L, et al. FLASH Irradiation Spares Lung Progenitor Cells and Limits the Incidence of Radio-induced Senescence [published online ahead of print 2019/12/05]. Clin Cancer Res. 2020;26(6):1497–1506. [DOI] [PubMed] [Google Scholar]
  • 42.Khan S, Bassenne M, Wang J, et al. Multicellular Spheroids as In Vitro Models of Oxygen Depletion During FLASH Irradiation [published online ahead of print 2021/02/06]. Int J Radiat Oncol Biol Phys. 2021. doi: 10.1016/j.ijrobp.2021.01.050. [DOI] [PubMed] [Google Scholar]
  • 43.Hendry JH, Moore JV, Hodgson BW, Keene JP. The constant low oxygen concentration in all the target cells for mouse tail radionecrosis [published online ahead of print 1982/10/01]. Radiat Res. 1982;92(1):172–181. [PubMed] [Google Scholar]
  • 44.Weiss H, Epp ER, Heslin JM, Ling CC, Santomasso A. Oxygen depletion in cells irradiated at ultra-high dose-rates and at conventional dose-rates [published online ahead of print 1974/07/01]. Int J Radiat Biol Relat Stud Phys Chem Med. 1974;26(1):17–29. [DOI] [PubMed] [Google Scholar]
  • 45.Cao X, Zhang R, Esipova TV, et al. Quantification of oxygen depletion during FLASH irradiation in vitro and in vivo. International Journal of Radiation Oncology* Biology* Physics. 2021. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Jansen J, Knoll J, Beyreuther E, et al. Does FLASH deplete Oxygen? Experimental Evaluation for Photons, Protons and Carbon Ions. Medical Physics. 2021. [DOI] [PubMed] [Google Scholar]
  • 47.Montay-Gruel P, Meziani L, Yakkala C, Vozenin MC. Expanding the therapeutic index of radiation therapy by normal tissue protection [published online ahead of print 2018/04/26]. Br J Radiol. 2019;92(1093):20180008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Montay-Gruel P, Acharya MM, Gonçalves Jorge P, et al. Hypofractionated FLASH-RT as an Effective Treatment against Glioblastoma that Reduces Neurocognitive Side Effects in Mice [published online ahead of print 2020/10/17]. Clin Cancer Res. 2021;27(3):775–784. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Franken NA, Rodermond HM, Stap J, Haveman J, Van Bree C. Clonogenic assay of cells in vitro. Nature protocols. 2006;1(5):2315–2319. [DOI] [PubMed] [Google Scholar]
  • 50.Geiger GA, Parker SE, Beothy AP, Tucker JA, Mullins MC, Kao GD. Zebrafish as a “biosensor”? Effects of ionizing radiation and amifostine on embryonic viability and development. Cancer research. 2006;66(16):8172–8181. [DOI] [PubMed] [Google Scholar]
  • 51.Vozenin MC, Hendry JH, Limoli CL. Biological Benefits of Ultra-high Dose Rate FLASH Radiotherapy: Sleeping Beauty Awoken [published online ahead of print 2019/04/24]. Clin Oncol (R Coll Radiol). 2019;31(7):407–415. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Gabriel F, Gippner P, Grosse E, et al. The Rossendorf radiation source ELBE and its FEL projects. Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms. 2000;161:1143–1147. [Google Scholar]
  • 53.Pawelke J, Brand M, Hans S, et al. Electron dose rate and oxygen depletion protect zebrafish embryos from radiation damage. Radiotherapy and Oncology. 2021;158:7–12. [DOI] [PubMed] [Google Scholar]
  • 54.Withers HR, Elkind MM. Microcolony Survival Assay for Cells of Mouse Intestinal Mucosa Exposed to Radiation. International Journal of Radiation Biology and Related Studies in Physics, Chemistry and Medicine. 1970;17(3):261–267. [DOI] [PubMed] [Google Scholar]
  • 55.Adrian G, Konradsson E, Lempart M, Bäck S, Ceberg C, Petersson K. The FLASH effect depends on oxygen concentration [published online ahead of print 2019/12/12]. Br J Radiol. 2020;93(1106):20190702. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Field S, Bewley D. Effects of dose-rate on the radiation response of rat skin. International Journal of Radiation Biology and Related Studies in Physics, Chemistry and Medicine. 1974;26(3):259–267. [DOI] [PubMed] [Google Scholar]
  • 57.Inada T, Nishio H, Amino S, Abe K, Saito K. High dose-rate dependence of early skin reaction in mouse. International Journal of Radiation Biology and Related Studies in Physics, Chemistry and Medicine. 1980;38(2):139–145. [DOI] [PubMed] [Google Scholar]
  • 58.Bourhis J, Sozzi WJ, Jorge PG, et al. Treatment of a first patient with FLASH-radiotherapy. Radiotherapy and oncology. 2019;139:18–22. [DOI] [PubMed] [Google Scholar]
  • 59.Venkatesulu BP, Sharma A, Pollard-Larkin JM, et al. Author Correction: Ultra high dose rate (35 Gy/sec) radiation does not spare the normal tissue in cardiac and splenic models of lymphopenia and gastrointestinal syndrome. Scientific reports. 2020;10(1):1–3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Alaghband Y, Cheeks SN, Allen BD, et al. Neuroprotection of radiosensitive juvenile mice by ultra-high dose rate flash irradiation. Cancers. 2020;12(6):1671. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61.Montay-Gruel P, Markarian M, Allen BD, et al. Ultra-High-Dose-Rate FLASH Irradiation Limits Reactive Gliosis in the Brain. Radiation Research. 2020;194(6):636–645. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Allen BD, Acharya MM, Montay-Gruel P, et al. Maintenance of Tight Junction Integrity in the Absence of Vascular Dilation in the Brain of Mice Exposed to Ultra-High-Dose-Rate FLASH Irradiation. Radiation research. 2020;194(6):625–635. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63.Lefaix JL, Mignot J. Effects of radiation on skin surface contour: A study of immediate effects in pig skin [published online ahead of print 1996/11/01]. Skin Res Technol. 1996;2(4):177–181. [DOI] [PubMed] [Google Scholar]
  • 64.Lefaix JL, Daburon F. Diagnosis of acute localized irradiation lesions: review of the French experimental experience [published online ahead of print 1998/09/30]. Health Phys. 1998;75(4):375–384. [DOI] [PubMed] [Google Scholar]
  • 65.Lefaix JL, Martin M, Tricaud Y, Daburon F. Muscular fibrosis induced after pig skin irradiation with single doses of 192Ir gamma-rays [published online ahead of print 1993/06/01]. Br J Radiol. 1993;66(786):537–544. [DOI] [PubMed] [Google Scholar]
  • 66.Rhodes-Feuillette A, Verola O, Lefaix JL, et al. The correlation of the interferon response with pathology and thermography studies in a pig model for the evaluation of local irradiation lesions [published online ahead of print 1986/01/01]. Br J Radiol Suppl. 1986;19:117–121. [PubMed] [Google Scholar]
  • 67.Nolan MW, Kent MS, Boss MK. Emerging Translational Opportunities in Comparative Oncology With Companion Canine Cancers: Radiation Oncology [published online ahead of print 2019/12/12]. Front Oncol. 2019;9:1291. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 68.Stankeová S, Crompton NE, Blattmann H, et al. Apoptotic response of irradiated T-lymphocytes. An epidemiologic study in canine radiotherapy patients [published online ahead of print 2003/11/08]. Strahlenther Onkol. 2003;179(11):779–786. [DOI] [PubMed] [Google Scholar]
  • 69.Dewhirst MW, Connor WG, Sim DA. Preliminary results of a phase III trial of spontaneous animal tumors to heat and/or radiation: early normal tissue response and tumor volume influence on initial response [published online ahead of print 1982/11/01]. Int J Radiat Oncol Biol Phys. 1982;8(11):1951–1961. [DOI] [PubMed] [Google Scholar]
  • 70.Rossmeisl JH. New treatment modalities for brain tumors in dogs and cats [published online ahead of print 2014/12/03]. Vet Clin North Am Small Anim Pract. 2014;44(6):1013–1038. [DOI] [PubMed] [Google Scholar]
  • 71.Konradsson E, Arendt ML, Jensen KB, et al. Establishment and initial experience of clinical FLASH radiotherapy in canine cancer patients. Frontiers in Oncology. 2021;11. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 72.Ferlay J, Laversanne M, Ervik M, et al. Global Cancer Observatory: Cancer Tomorrow. Lyon, France: International Agency for Research on Cancer. https://gco.iarc.fr/tomorrow. Accessed 6/1, 2021.
  • 73.Harrington KJ. Ultrahigh Dose-rate Radiotherapy: Next Steps for FLASH-RT [published online ahead of print 2018/08/11]. Clin Cancer Res. 2019;25(1):3–5. [DOI] [PubMed] [Google Scholar]
  • 74.Griffin RJ, Prise KM, McMahon SJ, Zhang X, Penagaricano J, Butterworth KT. History and current perspectives on the biological effects of high-dose spatial fractionation and high dose-rate approaches: GRID, Microbeam & FLASH radiotherapy [published online ahead of print 2020/07/25]. Br J Radiol. 2020;93(1113):20200217. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 75.Rahman M, Ashraf MR, Zhang R, et al. Electron FLASH delivery at treatment room isocenter for efficient reversible conversion of a clinical LINAC. International Journal of Radiation Oncology* Biology* Physics. 2021. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 76.Lempart M, Blad B, Adrian G, et al. Modifying a clinical linear accelerator for delivery of ultra-high dose rate irradiation. Radiotherapy and Oncology. 2019;139:40–45. [DOI] [PubMed] [Google Scholar]
  • 77.Montay-Gruel P, Bouchet A, Jaccard M, et al. X-rays can trigger the FLASH effect: Ultra-high dose-rate synchrotron light source prevents normal brain injury after whole brain irradiation in mice [published online ahead of print 2018/08/31]. Radiother Oncol. 2018;129(3):582–588. [DOI] [PubMed] [Google Scholar]
  • 78.Diffenderfer ES, Verginadis II, Kim MM, et al. Design, Implementation, and in Vivo Validation of a Novel Proton FLASH Radiation Therapy System. Int J Radiat Oncol Biol Phys. 2020;106(2):440–448. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 79.Patriarca A, Fouillade C, Auger M, et al. Experimental set-up for FLASH proton irradiation of small animals using a clinical system. International Journal of Radiation Oncology* Biology* Physics. 2018;102(3):619–626. [DOI] [PubMed] [Google Scholar]
  • 80.Wu YF, No HJ, Breitkreutz DY, et al. Technological Basis for Clinical Trials in FLASH Radiation Therapy: A Review. Appl Rad Oncol. 2021;10(2):6–14. [Google Scholar]
  • 81.DesRosiers C, Moskvin V, Bielajew AF, Papiez L. 150–250 MeV electron beams in radiation therapy. Physics in Medicine & Biology. 2000;45(7):1781. [DOI] [PubMed] [Google Scholar]
  • 82.Bazalova‐Carter M, Qu B, Palma B, et al. Treatment planning for radiotherapy with very high‐energy electron beams and comparison of VHEE and VMAT plans. Medical physics. 2015;42(5):2615–2625. [DOI] [PubMed] [Google Scholar]
  • 83.Schüler E, Eriksson K, Hynning E, et al. Very high‐energy electron (VHEE) beams in radiation therapy; Treatment plan comparison between VHEE, VMAT, and PPBS. Medical physics. 2017;44(6):2544–2555. [DOI] [PubMed] [Google Scholar]
  • 84.Palma B, Bazalova-Carter M, Hårdemark B, et al. Assessment of the quality of very high-energy electron radiotherapy planning. Radiotherapy and Oncology. 2016;119(1):154–158. [DOI] [PubMed] [Google Scholar]
  • 85.DesRosiers C, Moskvin V, Cao M, Joshi C, Langer M. Lung tumor treatment with very high energy electron beams of 150–250 Mev as compared to conventional megavoltage photon beams. International Journal of Radiation Oncology, Biology, Physics. 2008;72(1):S612. [Google Scholar]
  • 86.Papiez L, DesRosiers C, Moskvin V. Very high energy electrons (50–250 MeV) and radiation therapy. Technology in cancer research & treatment. 2002;1(2):105–110. [DOI] [PubMed] [Google Scholar]

RESOURCES