Comparison of the results obtained from the immunosensor with the actual additions.

Samples The concentration of S. aureus (CFU per mL) SD RSD (%)
Containing Added Immunosensor Plate count
1 ND 2.20 × 105 1.99 × 105 2.14 × 105 1.30 × 104 6.54
2 ND 2.20 × 106 2.04 × 106 2.07 × 106 1.09 × 105 5.32
3 ND 2.20 × 107 2.27 × 107 2.13 × 107 7.17 × 105 3.16
4 ND 2.20 × 108 2.11 × 108 2.10 × 108 8.60 × 106 3.91
5 ND 2.20 × 109 2.24 × 109 2.09 × 109 1.08 × 105 4.83