Published in final edited form as: J Magn Reson Imaging. 2021 Mar 19;54(4):1257–1265. doi: 10.1002/jmri.27608

TABLE 4.

Comparison of Other Imaging Artifact Burden on Late Gadolinium Enhancement (LGE) Sequences by Implantable Cardioverter-Defibrillator (ICD) Subtype

WB Other Artifact (%)
Standard LGE Other Artifact (%)
Median (IQR) N Median (IQR) N P-value
TV-ICD 0 (0–18)a,c 48 7 (0–21)d,f 47 0.43
CRT-D 23 (8–33)b,c 48 24 (10–33)e,f 14 0.73
S-ICD 0 (0–10)a,b 17 0 (0–7)d,e 7 0.71

As in Table 2, other non-device-related artifact (motion or wrap) is expressed as a percentage of short-axis slices affected by artifact out of the total number of short-axis slices.

TV-ICD = transvenous ICD; CRT-D = cardiac resynchronization therapy device; S-ICD = subcutaneous ICD; IQR = interquartile range; WB = wideband.

a

S-ICD versus TV-ICD for WB: P < 0.001.

b

S-ICD versus CRT-D for WB: P = 0.07.

c

TV-ICD versus CRT-D for WB: P = 0.10.

d

S-ICD versus TV-ICD for Standard LGE: P = 0.004.

e

S-ICD versus CRT-D for Standard LGE: P = 0.02.

f

TV-ICD versus CRT-D for Standard LGE: P = 0.19.