Skip to main content
. 2022 Apr 26;22:121. doi: 10.1186/s12871-022-01666-w

Table 2.

Airway management details

Management Details JEA
(n = 92)
LMA Supreme (n = 90) P
Implantation
 Time (s): 20.64 ± 0.40 20.62 ± 0.35 0.25
 No. attempts(1/2/3): 86/5/1 87/2/1 0.53
 one-time success rate: 93.48% 96.67% 0.50
Quality of ventilation (1/2/3) 89/3/0 88/2/1 0.55
Peak airway pressure(cm H2O) 16.87 ± 0.27 17.02 ± 0.25 0.68
FBS view of VC (optimal/suboptimal) 87/5 86/4 0.76

Data are expressed as number of patients or mean ± SD

FBS fiberscope, JEA Jcerity Endoscoper Airway, LMA laryngeal mask airway, VC vocal cords

*P ≤ 0.05 (JEA vs.LMA Supreme)