Effects of power-on time and ambient temperature on the output values from the colorimeter and the calculated absorbance (Mean ± standard deviation, n = 5).

Output value
Absorbance
I0 (Blank) I (0.25 mM) A = −log(I/I0)
Power-on time (at 25 °C) 0 h 405.2 ± 8.0 133.8 ± 4.9 0.481 ± 0.0081
0.5 h 403.2 ± 8.2 132.6 ± 5.2 0.483 ± 0.0085
2.0 h 404.0 ± 9.2 132.6 ± 5.2 0.484 ± 0.0073
Temperature 4 °C 371.4 ± 7.2 121.2 ± 4.6 0.487 ± 0.0081
40 °C 421.2 ± 8.3 139.2 ± 5.3 0.481 ± 0.0078