Table 3.

R-wave amplitudes and noise burden (Home Monitoring Service Center data)

R wave and noise by ICM position N= Mean ± SD Median (IQR)
R-wave amplitude, mV
 All insertion positions 621 0.73 ± 0.40 0.60 (0.42–0.97)
 Parallel to the heart's long axis 339 0.80 ± 0.43 0.71 (0.46–1.09)
 Parasternal position 259 0.62 ± 0.34 0.53 (0.38–0.77)
Noise burden, minutes/day
 All insertion positions 621 43.3 ± 102.4 9.4 (0.3–36.4)
 Parallel to the heart's long axis 339 36.3 ± 102.7 3.7 (0.0–26.8)
 Parasternal position 259 53.3 ± 104.8 14.5 (3.5–49.1)
R-wave amplitude stability: SD/mean, % 621 8.6 ± 6.5 7.0 (5.1–9.6)

ICM = insertable cardiac monitor; IQR = interquartile range; SD, standard deviation.

P < .001 (t test) and

P < .001 (Mann–Whitney U test).