Skip to main content
. 2022 Feb 13;3(2):169–175. doi: 10.1016/j.hroo.2022.02.005

Table 3.

Summary of device behavior at ERI in pacemakers by major manufacturer

Brand Behavior at ERI
Mode Rate Features disabled
Medtronic VVI Set rate (65 bpm) or programmed rate decrease (7%–10%) Hysteresis
Sleep function ventricular capture management
Atrial and ventricular sensing assurance
Vitatron VVI Set rate (65 bpm) or programmed rate decrease (10%–20%) or increased R-R interval (100 ms) EGM/episode/diagnostic data collection
Rate response
Post-PVC response
Flywheel mode
EGM range
PVC-synchronous Astim
Tachycardia fallback rate
AF prevention therapies
Ventricular rate stabilization
Therapy advisor
Boston Scientific No change at ERI; VVI at EOL No change at ERI; 50 bpm at EOL
Guidant VVI Set rate (65 bpm) or programmed rate decrease (11%–20%) Automatic capture
Rate response
St. Jude§ No change Increased R-R interval (100 ms) Rate sensor
NIPS test
AF suppression
Biotronik Single or dual chamber, depending on programmed mode Programmed rate decrease (4.5%–11%) Atrial pacing
Rate adaptation
Atrial and ventricular capture control
Atrial overdrive pacing
EGM recordings
Rate hysteresis
Night program
Rate fading
Statistics
Home monitoring
Ventricular pacing suppression
ELA Medical (MicroPort) VVI Set rate (70 bpm) Rate response
Smoothing
Rate hysteresis

AF = atrial fibrillation; EGM = electrogram; EOL = end of life; ERI = elective replacement indicator; NIPS = noninvasive programmed stimulation; PVC = premature ventricular contraction.

Medtronic Inc., Minneapolis, Minnesota, United States

Boston Scientific, Marlborough, Massachusetts, United States

Maastricht, The Netherlands

§

Abbott Laboratories, Abbott Park, Illinois, United States

Lake Oswego, Oregon, United States

Arvada, Colorado, United States