Table 2.
A comparative specification analysis of microwave with autoclave used in AIIMS, New Delhi
| S. no | Parameters | Microwave-assisted cold sterilization (MACS) | Autoclave (steam-based sterilization) |
|---|---|---|---|
| 1 | Basic Principle | Dielectric heating principle | Steam-based high-pressure principal |
| 2 | Mobility | Yes | No |
| 3 | Skilled Labor Required | No | Yes |
| 4 | Metal corrosion /contamination | No | Yes |
| 5 | Plastic fibre treatment | Yes | No |
| 6 | Internet hub connectivity | Yes | No |
| 7 | Process timing | 45 min | 120 min |
| 8 | Consumables | Nil | Pressure-sensitive tapes, bags, release effluent |
| 9 | Energy required | 3 Kw/h, 16 Amp, 1Ø | 6 Kw/h, 25 Amp, 3Ø |
| 10 | Chamber capacity | 500 L | 130 L |
| 11 | Water consumption | 800 ml per cycle | 130 L per cycle |
| 12 | Capex cost | Rs 24 lakhs | Rs 9 lakhs |