Skip to main content
. 2022 Mar 7;10(2):e00993-21. doi: 10.1128/spectrum.00993-21

TABLE 2.

Performance characteristics of RDT nAb test line based on nAb results determined by VMN assay (n = 112)

Performance characteristics Value 95% CI
Sensitivity 94.37% 86.20% to 98.44%
Specificity 92.68% 80.08% to 98.46%
Positive likelihood ratio 12.9 4.33 to 38.39
Negative likelihood ratio 0.06 0.02 to 0.16
Positive predictive value 95.71% 88.24% to 98.52%
Negative predictive value 90.48% 78.51% to 96.11%