Skip to main content
. 2022 Apr 21;2022:7339306. doi: 10.1155/2022/7339306

Table 2.

Interobserver reproducibility results for ETMs using REVO NX based on the first readings from each session taken by 2 operators in a sample of 40 eyes.

Sector Mean ± SD TRT Sw CoV (%) ICC
Superior (μm) 51.40 ± 3.31 2.63 0.95 1.85 0.92
Superior-temporal (μm) 51.38 ± 3.11 3.07 1.11 2.16 0.87
Temporal (μm) 51.38 ± 3.10 2.70 0.97 1.90 0.90
Inferior-temporal (μm) 51.38 ± 3.08 2.48 0.90 1.74 0.91
Inferior (μm) 51.30 ± 3.03 2.69 0.97 1.89 0.90
Inferior-nasal (μm) 51.32 ± 2.97 2.95 1.06 2.07 0.87
Nasal (μm) 51.32 ± 3.03 2.46 0.89 1.73 0.91
Superior-nasal (μm) 51.45 ± 2.94 2.97 1.07 2.08 0.86
Central (μm) 51.47 ± 3.25 2.00 0.72 1.40 0.95
Superior_Out (μm) 51.40 ± 2.81 3.78 1.36 2.65 0.76
Superior-temporal_Out (μm) 51.50 ± 2.70 3.59 1.30 2.52 0.77
Temporal_Out (μm) 51.38 ± 2.66 4.06 1.46 2.85 0.69
Inferior-temporal_Out (μm) 51.70 ± 2.53 3.36 1.21 2.34 0.77
Inferior_Out (μm) 51.50 ± 2.70 3.59 1.30 2.52 0.77
Inferior-nasal_Out (μm) 51.98 ± 2.66 3.36 1.21 2.33 0.79
Nasal_Out (μm) 51.50 ± 2.78 3.66 1.32 2.57 0.77
Superior-nasal_Out (μm) 51.48 ± 2.75 4.80 1.73 3.37 0.60