Summary of validation parameters of chromatographic method used for determination of simeprevir.

Parameter Simeprevir
Linearity
Regression equation Y = 54.75X + 15.28
Range (μg mL−1) 0.1–20
Correlation coefficient (r) 0.9997
Slope 54.75
Intercept 15.28
LOD (μg mL−1) 0.10
LOQ (μg mL−1) 0.34
Precision
Repeatability (intra-day) (% RSD) a
QCL (0.5 μg mL−1) 1.80%
QCM (7 μg mL−1) 1.03%
QCH (15 μg mL−1) 0.71%
Intermediate precision (inter-day) (% RSD) a
QCL (0.5 μg mL−1) 0.70%
QCM (7 μg mL−1) 0.44%
QCH (15 μg mL−1) 0.57%
Accuracy
(Mean ± S.D) b
QCL (0.5 μg mL−1) 100.80 ± 1.54
QCM (7 μg mL−1) 99.43 ± 1.34
QCH (15 μg mL−1) 100.67 ± 0.63
a

RSD: relative standard deviation.

b

Expressed mean of three replicates.