Skip to main content
. 2019 Jan 23;9(6):2997–3003. doi: 10.1039/c8ra09511f

Comparison of sensors for detection of AA, DA, and UA.

Electrode Method Linear range (μM) LOD (μM) Reference
AA DA UA AA DA UA
RGO-ZnO/GCE DPV 50–2350 1–70 3–330 3.71 0.33 1.08 6
AuNPs@MoS2/GCE DPV 1000–70 000 0.05–4000 10–7000 100 0.05 10 29
Fe-Meso-PANI/GCE LSVa 10–300 10–300 10–300 6.5 9.8 5.3 16
GO-PANI/GCE DPV 150–1050 1–14 3–26 50 0.5 1 46
Au/RGO/GCE DPV 240–1500 6.8–41 8.8–53 51 1.4 1.8 47
Graphene/SnO2/GCE DPV 100–1000 1–20 3–21 100 1 3 48
MoS2PANI/rGO/GCE DPV 50–8000 5.0–500 1.0–500 22.20 0.70 0.36 This work
a

LSV is linear sweep voltammetry.