Analytical parametersa.
LC | MS | |||
---|---|---|---|---|
RRLC | Agilent 1200 | MS | Agilent 6410 QQQ | |
Isocratic mobile phase | 45% ACN | ESI | Positive ESI | |
10 mM NH4COOH | 55% | Drying gas: N2 of low purity at 12 L min−1, pressure (60 psi) | ||
pH: 4.2 | ||||
Flow rate: 0.2 mL min−1 | ||||
Injection volume: 1 μL | ||||
Agilent Eclipse plus C18 column | 100 mm long | Source T: 350 °C | ||
2.1 mm internal diameter | Capillary V: 4000 V | |||
1.8 μm particle size | Collision cell | N2 (high purity) | ||
T: 20 ± 2 °C. | Mode | MRM | ||
Mass spectra segment | 0.0 to 1.2 min | Flow to waste | Analyte: naquotinib (NQT) | m/z 563 → m/z 463, FVb: 125 V, CEc: of 22 eV |
1.2 to 2.7 min | NQT MRM | m/z 563 → m/z 323, FV: 130 V, CE: of 18 eV | ||
2.7 to 4.0 min | FTB MRM | Foretinib (IS) | m/z 633 → m/z 128, FV: 145 V, CE: 20 eV |
LC, liquid chromatography; MS, mass spectrometry.
Fragmentor voltage.
Collision energy.