. 2019 Feb 7;9(9):4862–4869. doi: 10.1039/c8ra09812c

Analytical parametersa.

LC MS
RRLC Agilent 1200 MS Agilent 6410 QQQ
Isocratic mobile phase 45% ACN ESI Positive ESI
10 mM NH4COOH 55% Drying gas: N2 of low purity at 12 L min−1, pressure (60 psi)
pH: 4.2
Flow rate: 0.2 mL min−1
Injection volume: 1 μL
Agilent Eclipse plus C18 column 100 mm long Source T: 350 °C
2.1 mm internal diameter Capillary V: 4000 V
1.8 μm particle size Collision cell N2 (high purity)
T: 20 ± 2 °C. Mode MRM
Mass spectra segment 0.0 to 1.2 min Flow to waste Analyte: naquotinib (NQT) m/z 563 → m/z 463, FVb: 125 V, CEc: of 22 eV
1.2 to 2.7 min NQT MRM m/z 563 → m/z 323, FV: 130 V, CE: of 18 eV
2.7 to 4.0 min FTB MRM Foretinib (IS) m/z 633 → m/z 128, FV: 145 V, CE: 20 eV
a

LC, liquid chromatography; MS, mass spectrometry.

b

Fragmentor voltage.

c

Collision energy.