Skip to main content
. 2019 Feb 7;9(9):4862–4869. doi: 10.1039/c8ra09812c

Intra-day and inter-day (accuracy and precision) of established methoda.

HLM matrix LLQC (5 ng mL−1) LQC (15 ng mL−1) MQC (150 ng mL−1) HQC (400 ng mL−1)
Intra-day assayb Inter-day assayc Intra-day assay Inter-day assay Intra-day assay Inter-day assay Intra-day assay Inter-day assay
Mean 5.28 5.32 14.85 14.75 148.18 147.18 397.32 393.32
SD 0.08 0.05 0.23 0.38 2.49 2.71 2.76 6.70
Precision (%RSD) 1.52 0.99 1.52 2.58 1.68 1.84 0.70 1.70
% accuracy −5.65 −6.36 1.00 1.69 1.21 1.88 0.67 1.67
a

HLM, human liver microsome; LLQC, lower limit quality control; LQC, low quality control; MQC, medium quality control; HQC, high quality control.

b

Mean of twelve replicates on the same day.

c

Mean of six replicates for three days.