Statistical parameters and precision values of the proposed SWV method using AgNPs–rGO/PBB modified electrode.

Parameters The calculated value
Linearity range (mol L−1) 5 × 10−9 to 5 × 10−7
Intercept (μA) ± SDa 25.5 ± 2.34
Slope (μA (mol L−1)−1) ± SDb 3.64 × 109 ± 9.21 × 10−7
SSEc 50.1 × 10−8
Correlation coefficient (R) 0.9988
LODd (mol L−1) 2.17 × 10−9
LOQe (mol L−1) 6.66 × 10−9
Intra-day precisionf (% RSD, n = 3g) ≤3.45
Inter-day precisionf (% RSD, n = 3g) ≤3.96
a

Standard deviation of intercept.

b

Standard deviation of slope.

c

Sum of square errors.

d

Limit of detection.

e

Limit of quantitation.

f

Relative standard deviation.

g

Mean values of three measurements at three different concentration levels.