Skip to main content
. 2022 Apr 20;10:819534. doi: 10.3389/fcell.2022.819534

TABLE 2.

An overview of the imaging techniques used to study the structure and composition of cell-cell adhesions with electron and X-ray microscopy; TEM: transmission electron microscopy; FFEM: freeze-fracture electron microscopy; PREM: platinum replica electron microscopy; ET: electron tomography; VEM: volume electron microscopy; SBF-SEM: serial block face scanning electron microscopy; FIB-SEM: focused ion beam scanning electron microscopy; SXT: soft X-ray tomography; immuno-EM: immuno electron microscopy.

Imaging technique Resolution Sample thickness To investigate cell-cell adhesion and availability Cons In vivo imaging
TEM xy: 2 nm 60–80 nm Ultrastructure of the tissue at a nanometer resolution, morphology of junctions, relationship with the actin cytoskeleton, measuring intermembrane space Some technical handling and sample preparation optimization Not possible
Correlation with LM possible
Easily accessible and widely used
FFEM and PREM possible
ET xy: 2 nm Up to 300 nm Filament arrangement within junctions, binding interactions between substructures, quantitative measurements Some technical handling and sample preparation optimization Not possible
Easily accessible and widely used
VEM (including SBF-SEM and FIB-SEM) xy: up to 5 nm Up to 20 µm per day Junctions and ultrastructure in volume, quantification possible, relationship between different junctions and its localization on the membrane surface Technically demanding and sample preparation optimization necessary Not possible
z: up to 5 nm Limited availability, expensive machines
SXT xy: 10–50 nm 10 µm Junctions at a nanometer resolution in the near-native state and possible to reconstruct the ultrastructure of the cell in 3D; Correlation with LM possible Technically demanding and sample preparation optimization necessary Not possible
z: 10–50 nm Limited availability because need for synchrotron
Immuno-EM Same as TEM Same as TEM Individual proteins and the ultrastructure of the sample in a nanometer resolution Some technical handling and sample preparation optimization Not possible
Easily accessible and widely used