Table 2.
Postoperative LogMAR distance visual acuity, refractive status, and prediction error. Values provided are mean (standard deviation) and range. P value is from the Kruskall-Wallis test.
Healthy | Coexisting pathology | P | |
---|---|---|---|
Binocular uncorrected distance visual acuity | −0.06 ± 0.06 | −0.05 ± 0.06 | 0.604 |
0.00 to −0.20 | 0.10 to −0.20 | ||
| |||
Binocular corrected distance visual acuity | −0.06 ± 0.06 | −0.06 ± 0.06 | 0.887 |
0.00 to −0.20 | 0.00 to −0.20 | ||
| |||
Monocular uncorrected distance visual acuity | −0.01 ± 0.07 | 0.03 ± 0.08 | 0.027 |
0.30 to −0.20 | 0.30 to −0.10 | ||
| |||
Monocular corrected distance visual acuity | −0.01 ± 0.06 | 0.00 ± 0.06 | 0.300 |
0.20 to −0.20 | 0.20 to −0.10 | ||
| |||
Sphere (diopters) | −0.02 ± 0.08 | −0.06 ± 0.16 | 0.137 |
−0.50 to 0.00 | −0.75 to 0.00 | ||
| |||
Cylinder (diopters) | −0.06 ± 0.26 | −0.04 ± 0.16 | 0.993 |
−1.75 to 0.00 | −0–75 to 0.00 | ||
| |||
Spherical equivalent (diopters) | −0.04 ± 0.17 | −0.07 ± 0.19 | 0.229 |
−0.87 to 0.00 | −0.75 to 0.00 | ||
| |||
Mean prediction error (diopters) | 0.14 ± 0.20 | 0.06 ± 0.23 | 0.148 |
−0.56 to 0.54 | −0.52 to 0.43 | ||
| |||
Mean absolute prediction error (diopters) | 0.20 ± 0.14 | 0.20 ± 0.13 | 0.748 |
0.01 to 0.56 | 0.01 to 0.52 |