Recovery of glucose in human serum samples at the GCE/TiO2 NW/PAPBA–Au TNC electrode.
Sample | Added glucose [mM] | Concentration [mM] | Recovery [%] | RSDa [%] | |
---|---|---|---|---|---|
Determined by glucometer | Determined at the fabricated electrode | ||||
Human serum samples | 0 | 4.44 ± 0.01 | 4.43 ± 0.04 | 99.8 | |
1 | 5.415 ± 0.06 | 5.412 ± 0.01 | 99.94 | ||
1 | 5.44 ± 0.02 | 5.43 ± 0.05 | 99.81 | ||
1 | 5.43 ± 0.07 | 5.413 ± 0.04 | 99.68 | ||
2 | 6.43 ± 0.03 | 6.425 ± 0.07 | 99.92 | ||
2 | 6.36 ± 0.01 | 6.358 ± 0.02 | 99.97 | ||
2 | 6.383 ± 0.05 | 6.38 ± 0.03 | 99.95 | ||
3 | 7.388 ± 0.3 | 7.37 ± 0.01 | 99.75 | 2.84 | |
3 | 7.34 ± 0.03 | 7.35 ± 0.07 | 100.13 | ||
3 | 7.442 ± 0.04 | 7.31 ± 0.06 | 98.22 | ||
4 | 8.41 ± 0.01 | 8.43 ± 0.02 | 100.23 | ||
4 | 8.45 ± 0.06 | 8.442 ± 0.05 | 99.90 | ||
4 | 8.378 ± 0.02 | 8.4 ± 0.04 | 100.26 | ||
5 | 9.42 ± 0.07 | 9.431 ± 0.03 | 100.11 | ||
5 | 9.39 ± 0.03 | 9.388 ± 0.04 | 99.97 | ||
5 | 9.41 ± 0.05 | 9.414 ± 0.03 | 100.04 |
Number of measurements = 3.