. 2018 Mar 23;8(21):11517–11527. doi: 10.1039/c8ra00745d

Electrochemical response characteristics of the proposed SPE sensors.

Parameter Bare SPE Functionalized SPE Au NP SPE
Slope (mV per decade) −57.5 −58.3 −59.05
Intercept (mV) 229.33 235.8 249.05
Correlation coefficient 0.9994 0.9998 0.9996
LOD (M) 7.9 × 10−6 7.6 × 10−6 6.8 × 10−6
Response time (sec) 15 15 10
Isothermal coefficient (V °C−1) 0.00189 0.00187 0.00184
Working pH range 4–8 4–8 4–8
Concentration range (mol L−1) 10−5–10−2 10−5–10−2 10−5–10−2
Stability (months) 5 6 7
Accuracy (% R) 100.78 101.12 99.62
Precision (% RSD)
Repeatabilitya 1.345 0.987 0.659
Intermediate precisionb 1.453 0.876 0.531
Robustnessc (mean ± % RSD) 99.02 ± 1.026 98.45 ± 1.106 99.32 ± 0.872
a

Average of three different concentrations (1 × 10−4, 1 × 10−3 and 1 × 10−2 mol L−1) repeated three times within a day.

b

Average of three different concentrations (1 × 10−4, 1 × 10−3 and 1 × 10−2 mol L−1) repeated three times in three days.

c

Variation in method parameters such as the pH of the sample.