Skip to main content
. 2018 Mar 23;8(21):11517–11527. doi: 10.1039/c8ra00745d

Statistical comparison of the results obtained by the proposed SPE sensors and the official methoda.

Taken (mol L−1) Sensor 1 Sensor 2 Sensor 3 Official method
Found (mol L−1) % recovery Found (mol L−1) % recovery Found (mol L−1) % recovery Taken (μg mL−1) Found (μg L−1) % recovery
1 × 10−5 0.9823 × 10−5 98.23 0.9913 × 10−5 99.13 1.0023 × 10−5 100.23 2 1.991 99.55
5 × 10−5 5.057 × 10−5 101.14 4.942 × 10−5 98.84 4.917 × 10−5 98.34 4 4.0032 100.08
1 × 10−4 0.986 × 10−4 98.60 0.9945 × 10−4 99.45 0.99 × 10−4 99.00 6 5.887 98.12
5 × 10−4 4.988 × 10−4 99.76 4.982 × 10−4 99.64 4.9325 × 10−4 98.65 8 7.9086 98.87
1 × 10−3 1.0022 × 10−3 100.22 1.0087 × 10−3 100.87 1.011 × 10−3 101.10 10 9.909 99.09
Mean ± S.D. 99.59 ± 1.189 99.58 ± 0.780 99.46 ± 1.162 99.14 ± 0.736
% RSD 1.194 0.783 1.168 0.742
Variance 1.414 0.609 1.350 0.541
Student’s t-test 0.716 (2.306)b 0.925 (2.306)b 0.523 (2.306)b
F-value 2.610 (6.338)b 1.123 (6.338)b 2.493 (6.338)b
a

Official method (HPLC method using an octadecylsilane column and a methanol–0.01 mol L−1 ammonium phosphate buffer at pH 6.0 (60 : 40) as the mobile phase).

b

The values in parentheses are the corresponding tabulated t and F values at P = 0.05.