| 1 |
Image sensing device 340–380 nm |
Using image sensing device illumination of ultraviolet light in the wavelength range of 340–380 nm, fluorescence property of the different skin anomalies were detected |
Real time recording can be done and non-invasive |
Highly expensive |
— |
205
|
| 2 |
Infrared camera |
Infrared imaging takes place in different region of the tissues, which is caused due to micro calcification or calcification in the benign lesion |
High resolution |
Skin temperature varies for external noise that affects the result |
— |
206
|
| 3 |
Infrared thermometer |
Thermal sensor that detect the affected area in a thermal map and calibrated using the processor (30) and corresponding output signals taken over a predetermined range of ambient temperatures |
High resolution |
Cost is high |
— |
207
|
| 4 |
Infrared detectors |
Temperature distribution of the body surface is performed using infrared thermography which produces thermogram that further used for diagnosing procedure |
High resolution |
Emission from other surfaces affects the measurement accuracy |
— |
208
|
| 5 |
Amber radiance 1 T IR camera |
Differentiation of melanoma and benign cutaneous lesions is performed using infrared thermography and found the sensitivity and specificity for different depths of the lesion |
Real time recording and non-invasive |
Need optimal instrumentation setup |
Se: 39% Sp: 100% (0–5 mm lesions), Se: 58% Sp: 98% (lesions >5–15 mm), Se: 95% Sp: 100% (lesions >15–30 mm), Se: 78% Sp: 89% (lesions >30 mm) |
209
|
| 6 |
Infrared thermal camera operated at 3 μm −5 μm |
Thermal responses for healthy and malignant tissue were compared and concluded that increased metabolic activity of melanoma lesion can be detected using infrared imaging |
High resolution |
Camera has only 2% of accuracy |
— |
19
|
| 7 |
Infrared thermal camera operated at 14 000 nm |
Symptoms of illness are detected using embedded devices which can be in the form of watch, glass, camera etc. Communication interface is attached to the embedded device so that the captured data is transferred into the local server of the hospital or other healthcare station |
Automatically the captured data's are transmitted. |
Cost is high. |
— |
203
|
| 8 |
Infrared camera |
Thermal imaging for early stage detection of cutaneous melanoma is performed and resulting surface temperature oscillations were recorded using infrared camera |
Real time recording |
Lower rate transmission |
— |
210
|
| 9 |
Electrochemical sensor, optical sensor |
Biosensors are used to detect specific biological analyze later by converting a biological entity into an electrical signal and analyzed. Biosensor has vast potential in detecting the cancer cells |
Have more resolution comparing to other application sensor |
Specific chemical analyzer is random in nature so difficult to interpret |
— |
201
|
| 10 |
Nd:YAG laser at 532 nm |
The imaging sensor is use to capture the affected area which has 7.4 μm pixel pitch, composed of 1000 by 1000 imaging elements with a nanowire optical filter |
The signal to noise ratio is high and consume less power |
Cost is high since nanowire fabrication is implemented |
— |
204
|
| 11 |
Anon contact IR temperature sensor at 8–14 μm |
The emitted IR range extends from 800 nm to few hundred micrometers which are detected for the temperature variation |
Mapping of temperature variation is perfect |
Resolution is less |
— |
211
|
| 12 |
Infrared detectors |
The surface the sample is sensed with the small change in the motion. This motion is detected by bending of cantilever in the sensor system. Such system provides information about the position of the subject and the temperature value during displacement |
Nanoparticles are synthesized in the sensor which makes the accuracy to be good |
Cost is high |
— |
212
|
| 13 |
IR absorbance sensor at 3300 nm and 3570 nm |
Infrared absorbance sensor used to detect the malignant cells in the biopsy. It takes 8 seconds to identify whether it is tumor or normal cell |
This sensor is used in diagnosing the melanoma during histopathological analysis and also decreases the wrong interpretation |
Supports shorter range and its performance degrade with longer distances |
— |
213
|