. 2022 May 5;15:121–129. doi: 10.2147/MDER.S366053

Table 1.

Evaluation of Our Device with Vital Sign Simulator and Standard Pulse Oximeter Device

Iteration Number Test on Prosim 8 Vital Sign Simulator Comparison with Standard Device
SPO2 Test Heart Rate Test SPO2 Test
Values from the Simulator Measurement by Our Prototype % Error Values from the Simulator Measurement by Our Prototype % Error Measurement by Standard Pulse Oximeter Measurement by Our Prototype % Error
1 70 67 4.29 120 123 2.50 68 67 1.47
2 72 69 4.17 115 117 1.74 70 69 1.43
3 74 71 4.07 110 113 2.73 73 71 2.74
4 76 74 2.63 105 107 1.90 75 74 1.33
5 78 76 2.56 100 103 3.00 77 76 1.30
6 80 78 2.5 95 98 3.16 79 78 1.27
7 81 79 2.47 90 92 2.22 80 79 1.25
8 83 81 2.41 85 86 1.18 83 81 2.41
9 85 83 2.35 80 80 0.00 84 83 1.19
10 88 86 2.27 75 76 1.33 88 86 2.27
11 90 89 1.11 70 71 1.43 90 89 1.11
12 93 92 1.08 65 66 1.54 93 92 1.08
13 95 96 1.05 60 62 3.33 96 96 0.00
14 97 97 0 55 52 5.45 98 97 1.02
15 99 98 1.01 50 46 8.00 100 98 2.00
Average % Error 2.26 2.63 1.46