Table 2.
The gamma passing rates at 1%/1 mm criteria of standard plans irradiation with different amplitudes and unpaired t-test statistical analysis between 1D (X or Y-axis) and 2D (XY-axis) moving patterns
| Amplitude (cm) | Moving platform in X-axis | Moving platform in Y-axis | Moving platform in XY-axis | Unpaired t-test (P) | |
|---|---|---|---|---|---|
|
| |||||
| X versus XY | Y versus XY | ||||
| 0.0 | 100.0 | 100.0 | 100.0 | - | - |
| 0.5 | 86.1 | 88.8 | 72.4 | 0.72 | 0.93 |
| 1.0 | 55.8 | 60.5 | 35.2 | 0.31 | 0.25 |
| 2.0 | 19.2 | 24.5 | 12.3 | 0.27 | 0.10 |